Xlumena, Inc.: Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures.
- Recall number
- Z-2167-2013
- Recalling firm
- Xlumena, Inc.
- Product
- Xlumena brand NAVIX Access Device, Model/Catalogue Number: NVX-10-03, Part Number: FG-00666, Product is manufactured and distributed by Xlumena, Inc., Mountain View, CA The NAVIX Access Device is a multi-lumen balloon catheter with a trocar, anchor balloon, and dilation balloon that creates and dilates a tract across tissue planes. The NAVIX Access Device is indicated for use as an accessory to cannulate the transgastric or transduodenal wall and into a pancreatic pseudocyst, when it is visibly bulging into the gastrointestinal tract during endoscopic procedures.
- Status
- Terminated
- Initiated
- 2013-04-17
- Posted
- 2013-09-09
- Terminated
- 2013-09-23
- Reason
- A complaint investigation has found that product fractures can occur at the distal end of the catheter under load.
- Root cause
- Process design
- Action
- Xlumena sent an Urgent Medical Device Recall letter dated April 17, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discontinue use and segregate the affected product. Customers with questions were instructed to contact dhasker@xlumena.com. For question regarding this recall call 650-961-9901.
- Quantity
- 148 units
- Distribution
- Worldwide Distribution - USA including AZ, CA, FL, GA, IN, NY,SC, WA, PA, IL, CO, and internationally to Netherlands, Germany, Spain, Denmark, and Switzerland.
- Lot, serial or model codes
- All lots are involved are being recalled, including FGS00048, FGS0005, FGS00058, FGS00059, FGS00069, FGS00073, FGS00081, FGS00086, FGS00091
- 510(k) numbers
- ["K093279"]
- PMA numbers
- not on file
- Product code
- GCJ
- Firm FEI
- 3008516478
- Firm city
- Mountain View
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28