Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc. Endoscopy Division: Smith & Nephew 5.5 mm TWINFIX Ultra Ti Anchor with two #2 sutures (COBRAID-blue, COBRAIDblack), sterile Part Number: 72202896 Product Usage: intended for use for the reattachment of soft tissue to bone

Recall number
Z-2167-2012
Recalling firm
Smith & Nephew, Inc. Endoscopy Division
Product
Smith & Nephew 5.5 mm TWINFIX Ultra Ti Anchor with two #2 sutures (COBRAID-blue, COBRAIDblack), sterile Part Number: 72202896 Product Usage: intended for use for the reattachment of soft tissue to bone
Status
Terminated
Initiated
2012-07-03
Posted
2012-08-09
Terminated
2016-04-18
Reason
Distal part of the anchor may break on insertion into bone during surgery
Root cause
Device Design
Action
Smith & Nephew sent an Urgent Product Recall 1st Notification letters dated June 2012 via Federal Express to all affected customers. The letter identified the affected product, potential risk and actions to be taken. Customers were instructed to inspect inventory and quarantine them immediately, complete and return the attached form indicating the Lot# and quantities that need to be returned and contact Smith & Nephew Returns Group at 800-343-5717 (option 3) or email to endo.andreturns@smith-nephew.com. for instructions for returning affected products and receiving a credit. For questions call 1-508-261-3731.
Quantity
2,204 units
Distribution
Worldwide Distribution - US Nationwide and the countries of: Argentina, AUSTRALIA, AUSTRIA, Belgium, CANADA,CHILE COLOMBIA, Denmark, DUBAI, FINLAND, FRANCE GDC BAAR Distribution,GERMANY, GREECE, INDIA ITALY, KOREA, MALAYSIA, MEXICO, NETHERLAND, NORWAY PORTUGAL, PUERTO RICO, SHANGHAI, SOUTH AFRICA SPAIN, SWEDEN, SWITZERLAND, and UNITED KINGDOM.
Lot, serial or model codes
Lot Numbers: 50388182, 50389253, 50391626, 50392272, 50392407, 50395081, 50395680, 50396197, 50400556, 50401558, 50402228, 50403940, 50406036, 50407582, 50409406, 50413407, 50414970, 50412647, 50415864, 50416801, 50418088, 50420654, 50421030, 50421495
510(k) numbers
["K100159"]
PMA numbers
not on file
Product code
MBI
Firm FEI
3003604053
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register