Skip to the content

FDA Medical Device Recalls (openFDA)

INNOVATIVE TOMOGRAPHY PRODUCT GMBH: Semi-automatic Biopsy-Needle, ITP innotom.com, CE 0297, STERILE EO for the following article numbers: Article no. BIM 18/10, Length 100 mm, Diameter 18 G (1.25 mm); Article no. BIM 18/15, Length 150 mm, Diameter 18 G (1.25 mm); Article no. BIM 18/20, Length 200 mm, Diameter 18 G (1.25 mm); Article no. BIM 16/10, Length 100 mm, Diameter 16 G (1.6 mm); Article no. BIM 16/15, Length 150 mm, Diameter 16 G (1.6 mm); Article no. BIM 16/20, Length 200 mm, Diameter 16 G (1.6 mm); Article no. BIM 14/10, Length 100 mm, Diameter 14 G (2.1 mm); Article no. BIM 14/15, Length 150 mm, Diameter 14 G (2.1 mm); Article no. BIM 14/20, Length 200 mm, Diameter 14 G (2.1 mm) *NOT DISTRIBUTED WITHIN THE US

Recall number
Z-2164-2021
Recalling firm
INNOVATIVE TOMOGRAPHY PRODUCT GMBH
Product
Semi-automatic Biopsy-Needle, ITP innotom.com, CE 0297, STERILE EO for the following article numbers: Article no. BIM 18/10, Length 100 mm, Diameter 18 G (1.25 mm); Article no. BIM 18/15, Length 150 mm, Diameter 18 G (1.25 mm); Article no. BIM 18/20, Length 200 mm, Diameter 18 G (1.25 mm); Article no. BIM 16/10, Length 100 mm, Diameter 16 G (1.6 mm); Article no. BIM 16/15, Length 150 mm, Diameter 16 G (1.6 mm); Article no. BIM 16/20, Length 200 mm, Diameter 16 G (1.6 mm); Article no. BIM 14/10, Length 100 mm, Diameter 14 G (2.1 mm); Article no. BIM 14/15, Length 150 mm, Diameter 14 G (2.1 mm); Article no. BIM 14/20, Length 200 mm, Diameter 14 G (2.1 mm) *NOT DISTRIBUTED WITHIN THE US
Status
Open, Classified
Initiated
2021-04-21
Posted
not on file
Terminated
not on file
Reason
Sterility Assurance; Coaxial Interventional Needle labeled as sterile may not have been adequately sterilized
Root cause
Vendor change control
Action
On April 20, 2021, Innovative Tomography Products GmbH issued a "urgent Safety Information Product Recall" notification to affected consignees via E-Mail. In addition to informing consignees about the recall products, the firm asked consignees mo to take the following actions: 1. We request that you immediately check your stock of the products in question with the LOT numbers indicated. Please block the corresponding products and quarantine them. 2. Please complete the attached response letter and return it to ITP via E-Mail to the following address: Address: ITP GmbH Rckrufunterlagen Universittsstrae 136 44799 Bochum Germany Contact person: Dominik Frmke Tel.: +49 234 54621942 Fax: +49 234 54622444 E-Mail: fraemke@innotom.com 3. Upon receipt of the documents, we will contact you to retrieve and exchange the products. Information sharing: 4. Please inform all persons within your company belonging to the circle of users about this urgent safety
Quantity
not on file
Distribution
US Nationwide distribution in the state of CA.
Lot, serial or model codes
Affected Lot Numbers:  BIM 18/10: LOT: 4918;  BIM 18/15: LOT: 4017, 4918, 4719;  BIM 18/20: LOT: 4017, 4918;  BIM 16/10: LOT: 4918, 1119, 1219;  BIM 16/15: LOT: 4918, 1119;  BIM 16/20: LOT: 4918, 1119;  BIM 14/10: LOT: 4918, 3019;  BIM 14/15: LOT: 4918;  BIM 14/20: LOT: 4918
510(k) numbers
not on file
PMA numbers
not on file
Product code
GAA
Firm FEI
3014408997
Firm city
Bochum
Firm state
not on file
Firm country
Germany

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register