GE Healthcare, LLC: Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthetic agent desflurane. The Tec 6 Plus attaches only to Datex-Ohmeda anesthesia machines equipped with the proprietary Selectatec manifold. The Tec 6 Plus NAD Variant attaches only to NAD anesthesia machines equipped with the NAD exclusion type manifold.

Recall number
Z-2161-2013
Recalling firm
GE Healthcare, LLC
Product
Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthetic agent desflurane. The Tec 6 Plus attaches only to Datex-Ohmeda anesthesia machines equipped with the proprietary Selectatec manifold. The Tec 6 Plus NAD Variant attaches only to NAD anesthesia machines equipped with the NAD exclusion type manifold.
Status
Terminated
Initiated
2013-07-29
Posted
2013-09-06
Terminated
2015-04-17
Reason
GE Healthcare has recently become aware of a potential safety issue involving the pre-operative check, low pressure leak test at 1% dial setting of the Tec 6 and Tec 6 Plus Vaporizers. The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to
Root cause
Component change control
Action
GE Healthcare sent an Urgent Medical Device Correction letter dated August 1, 2013, to all affected customers. The letter described the issue, asked that the Tec 6 and Tec 6 Plus Vaporizers user manual be updated with the attached addendum. Also Users are to perform the user manual recommended pre-operative check which includes the low pressure leak test on every vaporizer with the dial turned to the 12% setting instead of the specified 1% setting as per the attached addendum. If the device does not pass the new low pressure leak at 12%, it should be removed from service and consignee should ontact a GE Healthcare Representative for a repair/replacement. If the vaporizer passes the new low pressure leak test at the 12 % setting, it can continue to be used, however, as detailed in the user manual, continue to routinely perform the pre-operative check which includes the low pressure leak test.A Medical Device Correction Confirmation form was asked to be sent back to GE. Customers with questions or concerns were instructed to call 1-855-700-4350 or send an email to vaporizer.admin@ge.com. For countries outside the US, customers were instructed to contact their local GE Healthcare Service Representative. For questions regarding this recall call 262-513-4122.
Quantity
31,443 devices (23,947 US, 7496 OUS)
Distribution
Worldwide Distribution - USA (nationwide) Puerto Rico and internationally to AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CROATIA local name-Hrvatska, CYPRUS, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JORDAN, KOREA-REPUBLIC OF, KUWAIT, LITHUANIA, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, PORTUGAL, QATAR, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, and VENEZUELA,
Lot, serial or model codes
Part Numbers: X1107-9001-000, X1107-9002-000, X1107-9003-000, X1107-9004-000, X1107-9005-000, X1107-9006-000, X1107-9007-000, X1109-9008-000, X1109-9009-000, X1107-9010-000, X1107-9011-000, X1107-9012-000, X1107-9014-000, X1107-9015-000, X1107-9016-000, X1107-9101-000, X1107-9102-000, X1107-9103-000, X1107-9201-000, X1107-9202-000, X1107-9203-000, X1107-9205-000, X1107-9206-000, X1107-9601-000, X1107-9602-000, X1107-9603-000, X1107-9604-000, X1107-9605-000, X1107-9606-000, X1107-9607-000, X1107-9608-000, X1107-9609-000, X1107-9610-000, X1107-9611-000, X1107-9612-000, X1107-9614-000, X1107-9615-000, X1107-9616-000, X1107-9617-000, M1145610-X, M1220023-X, M1220854-X, M1221495-X, 2064455-X, M1243207-X, X1107-9701-000, X1107-9702-000, X1107-9703-000, X1107-9211-000, X1107-9212-000, X1107-9213-000, X1107-9215-000, X1107-9216-000, X1107-9217-000, X1107-9207-000, X1107-9208-000, 2066725-X, 2066835-X, 2069767-X, 1107-9601-000, 1107-9612-000, M1221495, M1220023, M1243207, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9605-000, 1107-9606-000, 1107-9607-000, 1107-9608-000, 1107-9609-000, 1107-9610-000, 1107-9611-000, 1107-9614-000, 1107-9615-000, 1107-9616-000, 1107-9617-000, M1145610, M1220854, 2064455-001, M1098876, 2066725-001, 2066835-001, 2069767-001, X1107-8003-000, X1107-8006-000, X1107-8007-000, X1107-8008-000, X1107-8012-000, X1107-8013-000, X1107-8014-000, X1107-8015-000, X1107-8016-000, X1107-8017-000, X1107-8018-000, X1107-8019-000
510(k) numbers
["K000275","K925580"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register