Howmedica Osteonics Corp.: Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100

Recall number
Z-2160-2019
Recalling firm
Howmedica Osteonics Corp.
Product
Serrato 9.5mm X 90mm Polyaxial Screw Catalog Number 482619590 Serrato 9.5mm X 100mm Polyaxial Screw Catalog Number 4826195100
Status
Terminated
Initiated
2019-06-18
Posted
not on file
Terminated
2020-05-22
Reason
Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.
Root cause
Labeling mix-ups
Action
1. Please note this product has already been removed by your local Stryker Sales Representative and/or Stryker Branch or Agency. This communication is being made for notification purposes only; there are no additional product return requirements for this matter. 2. Please assist us in meeting our regulatory obligation by completing and signing the enclosed Product Recall Acknowledgment Form and email a copy to Stryker Spine at SpineRegulatoryActions@Stryker.com or fax the form to 201-575-4675. If you have any questions, please contact Christa Joisil or Matthew Kelleher from the Regulatory Compliance Team: Spine-RegulatoryActions@Stryker.com or 201.749.8090
Quantity
11
Distribution
US Nationwide Distribution - CA, FL, GA, MA, NY, PA, TX
Lot, serial or model codes
Lot # B88751 UDI: (01)07613327351576(11)190125(10)B88751
510(k) numbers
["K170496"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register