Skip to the content

FDA Medical Device Recalls (openFDA)

ARROW INTERNATIONAL, LLC: Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM

Recall number
Z-2158-2026
Recalling firm
ARROW INTERNATIONAL, LLC
Product
Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM
Status
Open, Classified
Initiated
2026-04-10
Posted
2026-05-13
Terminated
not on file
Reason
16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
Root cause
Nonconforming Material/Component
Action
On 4/10/2026, recall notices were mailed to customers who were informed of the following Arrow International LLC is recalling product due to the Merit Medical Systems, Inc. recall of the Merit 16F Dual-Valved Splitable Sheath Introducer component, which is included in impacted Arrow Hemodialysis Kits & Sets. Customers were asked to do the following: 1) Users should cease use and distribution of affected product and immediately quarantine the affected product. 2) Complete and return the Acknowledgement Form, see link: https://fca.myteleflex.com/en/recall/000605 3) This notice should be passed on to all persons who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. In addition, distributors were asked to do the following: - Provide a copy of this field safety notice to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. Should you require further information or support concerning this issue, please contact Customer Service via email (Recalls@teleflex.com), phone (1-866-396-2111), or FAX (1-855-419-8507).
Quantity
34,568
Distribution
Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW
Lot, serial or model codes
REF/UDI-DI/Lots: CS-15242-VSP/30801902195273/33F23C0870, 33F23D0857, 33F23F0719, 33F23K0182, 33F24A0265, 33F24C0062, 33F24D1000, 33F24D1039, 33F24E0803, 33F24H0038, 33F24H0261, 33F24K0505, 33F24M0039, 33F24M0218, 33F25A0141, 33F25C0657, 33F25D0125, 33F25E0305, 33F25E0429, 33F25E0430, 33F25E1131, 33F25F0042, 33F25G0070, 33F25H0295, 33F25H0366, 33F25H0674, 33F25J1015; CS-15282-VSP/30801902195518/33F23C0866, 33F23D0854, 33F23E0572, 33F23F0594, 33F23G0540, 33F23J0544, 33F23K0174, 33F23L0950, 33F24B0010, 33F24C0060, 33F24D0993, 33F24D1040, 33F24E0268, 33F24E0809, 33F24F0155, 33F24J0037, 33F24M0007, 33F24M0040, 33F24M0165, 33F24M0217, 33F25A0406, 33F25B0269, 33F25C0021, 33F25C0045, 33F25C0642, 33F25D0013, 33F25D0713, 33F25F0249, 33F25H0296, 33F25J0051; CS-15322-VSP/20801902195184/33F23C0869, 33F23D0862, 33F23F0723, 33F23H0985, 33F24A0272, 33F24C0066, 33F24E0813, 33F24G0170, 33F24G0185, 33F24H0256, 33F24J0040, 33F24L0183, 33F24M0222, 33F25A0426, 33F25C0022, 33F25C0593, 33F25E1082, 33F25G0024, 33F25H0297; CS-15362-VSP/30801902195525/33F23C0875, 33F23G0546, 33F23K0195, 33F24E0186, 33F24H0039, 33F24L0201, 33F24M0235, 33F25C0647, 33F25E0326, 33F25G0058, 33F25J0043; CS-15552-VSP/30801902195198/33F23G0552, 33F23H0987, 33F23L0961, 33F24C0070, 33F24E0820, 33F24H0257, 33F24L0068, 33F24M0227, 33F25C0654, 33F25E0432; CSD-15242-SP/20801902096610/33F24A0303, 33F25C0688; CSD-15282-SP/20801902096634/33F23K1087, 33F24A0297; CS-15242-SPM/10801902193527/33F23D0867, 33F23K0192, 33F24A0288, 33F24E0802, 33F24M0244, 33F25C0669, 33F25E0338, 33F25G0586; CS-15282-SPM/10801902193688/33F23C0872, 33F23D0864, 33F23G0539, 33F23L0949, 33F24A0278, 33F24C0075, 33F24E0808, 33F24H0254, 33F24J0036, 33F24L1166, 33F24M0233, 33F25A0413, 33F25C0666, 33F25H0824; CS-15322-SPM/30801902193538/33F23D0865, 33F23H1010, 33F23L0955, 33F24E0812, 33F25C0682, 33F25E0335; CS-15362-SPM/20801902193548/33F24G0062
510(k) numbers
["K111117"]
PMA numbers
not on file
Product code
MSD
Firm FEI
3013162291
Firm city
Morrisville
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.