Beckman Coulter Inc.: UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858.
- Recall number
- Z-2158-2017
- Recalling firm
- Beckman Coulter Inc.
- Product
- UniCel DxH 600 Coulter Cellular Analysis System, Catalog No. B23858.
- Status
- Terminated
- Initiated
- 2017-04-24
- Posted
- not on file
- Terminated
- 2018-05-22
- Reason
- There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result.
- Root cause
- Device Design
- Action
- An Urgent Medical Device Recall letter dated 4/24/17 was sent to customers to inform them that there is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual light scatter pattern may be incorrectly removed from analysis which can result in an erroneous differential result. The letter informs the customers of the issue, impact, action, and resolution. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with any questions regarding the notice are instructed to contact Customer Support Center at Http://www.beckmancoulter.com/customersupport/support, 800-526-7694 in the US and Canada, customers outside the U.S. and Canada are instructed to contact their local Beckman Coulter representative.
- Quantity
- 1408 units total (998 units in US)
- Distribution
- Worldwide distribution: US nationwide, Algeria, Andorra, Austria, Australia, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Canada, China, Colombia, Costa Rica, Cote d'lvoire, Croatia, Czech Republic, Ecuador, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan , Kuwait, Lebanon, Lithuania ,Macao, Malaysia, Mayotte, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania , Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia , South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Republic, United Kingdom, Uruguay and Vietnam.
- Lot, serial or model codes
- all software versions
- 510(k) numbers
- ["K081930","K120771","K140911"]
- PMA numbers
- not on file
- Product code
- GKZ
- Firm FEI
- 2050012
- Firm city
- Brea
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28