Skip to the content

FDA Medical Device Recalls (openFDA)

American Medical Systems Innovation Center - Silicon Valley: Sureflex 273 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection and incision of soft and cartilaginous tissue and surgical procedures involving vaporization, ablation and fragmentation of calculi (urinary and biliary).

Recall number
Z-2158-2015
Recalling firm
American Medical Systems Innovation Center - Silicon Valley
Product
Sureflex 273 Lithotripsy Fiber, 5x, Blue; Sure F!ex laser fibers are intended for use in laser-based surgical applications, including, but not limited to endoscopic, laparoscopic and open surgical procedures involving vaporization, ablation, coagulation, hemostasis, excision, resection and incision of soft and cartilaginous tissue and surgical procedures involving vaporization, ablation and fragmentation of calculi (urinary and biliary).
Status
Terminated
Initiated
2015-05-28
Posted
2015-07-21
Terminated
2016-01-19
Reason
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex'" Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
Root cause
Reprocessing Controls
Action
Urgent Medical Device Field notification letters were sent on June 9, 2015. Translation for countries out of the US will be conducted. Communications are being coordinated by the Prostate Health QA Director. Urgent Medical Device Field notification Letters advised customers of the issue and list all affected products with name and model numbers. Any SureFlex Reuseable Fibers which have not been used but are in inventory and with the 3 year shelf life should be returned as per instructions enclosed. Any of the SureFlex devices in inventory that have been used should be discarded or returned. All Reuseable Stripper and Cleaver Accessories that have not been used should be returned. Used devices should be destroyed or returned. Questions to 1-408-456-3320. The Acknowledgement form should be completed and returned.
Quantity
~16,710 all affected device
Distribution
Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
Lot, serial or model codes
Model R-LLF273TG All product manufactured from April 8, 2012 to April 8, 2015
510(k) numbers
["K050108"]
PMA numbers
not on file
Product code
GEX
Firm FEI
2937094
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register