FDA Medical Device Recalls (openFDA)
Medtronic Powered Surgical Solutions: Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.
- Recall number
- Z-2156-2008
- Recalling firm
- Medtronic Powered Surgical Solutions
- Product
- Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.
- Status
- Terminated
- Initiated
- 2008-05-30
- Posted
- 2008-08-23
- Terminated
- 2009-08-22
- Reason
- Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.
- Root cause
- Process control
- Action
- Firm notified consignees by Urgent: Medical Device Recall letter sent May 30, 2008. The letter requires consignees to immediately compare their inventory of products against the list provided and quarantine any unused dissecting tools. The letter also states that Medtronic would provide replacement product to match the quantity and type returned by the consignees' institution.
- Quantity
- 1,306 units
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Lot numbers: 0000710253, 0000710254, 0000710256, 0000710257, 0000710258, 0000710266, 0000710267, 0000710271, 0000710272, 0000710274, 0000710275, 0000710276, 0000710277, 0000710278, 0000710279, 0000710280, 0000710281, and 0000710288
- 510(k) numbers
- ["K020069"]
- PMA numbers
- not on file
- Product code
- HBB
- Firm FEI
- 1625507
- Firm city
- Fort Worth
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as