Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Powered Surgical Solutions: Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.

Recall number
Z-2156-2008
Recalling firm
Medtronic Powered Surgical Solutions
Product
Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General Surgical and Plastic Surgery applications for transection, osteotomy, graft harvesting, bone shaping, laminectomy, entry hole, suture hole, and midface advancement.
Status
Terminated
Initiated
2008-05-30
Posted
2008-08-23
Terminated
2009-08-22
Reason
Cutting flute geometry of surgical dissecting tools may be outside of established tolerance. This may lead to reduced cutting performance, inability to perform desired dissection, increased risk of overheating, and potential need for revision surgery.
Root cause
Process control
Action
Firm notified consignees by Urgent: Medical Device Recall letter sent May 30, 2008. The letter requires consignees to immediately compare their inventory of products against the list provided and quarantine any unused dissecting tools. The letter also states that Medtronic would provide replacement product to match the quantity and type returned by the consignees' institution.
Quantity
1,306 units
Distribution
Nationwide Distribution.
Lot, serial or model codes
Lot numbers: 0000710253, 0000710254, 0000710256, 0000710257, 0000710258, 0000710266, 0000710267, 0000710271, 0000710272, 0000710274, 0000710275, 0000710276, 0000710277, 0000710278, 0000710279, 0000710280, 0000710281, and 0000710288
510(k) numbers
["K020069"]
PMA numbers
not on file
Product code
HBB
Firm FEI
1625507
Firm city
Fort Worth
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register