Skip to the content

FDA Medical Device Recalls (openFDA)

MEDLINE INDUSTRIES, LP - Northfield: Medline Convenience Kits, containing BD ChloraPrep Clear - 1mL Applicator, labeled as: 1) BLOOD CULTURE KIT, REF DYNDH1462B; 2) BLOOD CULTURE KIT, REF DYNDH1622C; 3) BLOOD CULTR KT DIVRSN DEVICE, REF DYNDH1866.

Recall number
Z-2155-2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Convenience Kits, containing BD ChloraPrep Clear - 1mL Applicator, labeled as: 1) BLOOD CULTURE KIT, REF DYNDH1462B; 2) BLOOD CULTURE KIT, REF DYNDH1622C; 3) BLOOD CULTR KT DIVRSN DEVICE, REF DYNDH1866.
Status
Open, Classified
Initiated
2025-06-12
Posted
2025-07-21
Terminated
not on file
Reason
Affected convenience kits contain BD ChloraPrep Clear - 1 mL Applicators, which were recalled by BD due to potential to exhibit an open seal on its packaging. This may constitute a breach to sterility, which may in turn lead to the infection related harms for the patient.
Root cause
Other
Action
On June 12, 2025, firm instructed customers to request stickers to over-label the affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No product is to be returned.
Quantity
15700 units
Distribution
Distributed domestically to California, Florida, Pennsylvania, and Texas. No international distribution.
Lot, serial or model codes
1) REF DYNDH1462B: UDI/DI 10195327662462 (EA), 40195327662463 (CASE), Lot Numbers: 24EBU164, 24FBD281, 24FBK195; 2) REF DYNDH1622C: UDI/DI 10195327509873 (EA), 40195327509874 (CASE), Lot Numbers: 24EBU004; 3) REF DYNDH1866: UDI/DI 10195327400958 (EA), 40195327400959 (CASE), Lot Numbers: 24EBT285.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OIB
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register