Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Electronics North America Corporation: Philips DigitalDiagnost 3.1.x X-Ray System

Recall number
Z-2154-2017
Recalling firm
Philips Electronics North America Corporation
Product
Philips DigitalDiagnost 3.1.x X-Ray System
Status
Terminated
Initiated
2017-05-11
Posted
not on file
Terminated
2018-07-10
Reason
During recent evaluations of the Philips DigitalDiagnost 3.1.x, we have identified a potential issue that may affect the performance of the equipment under certain conditions.
Root cause
Software design
Action
Philip's planned action to bring defect into compliance: 1. You will contact customers and initiate a software update to correct the issue that incorrectly notifies the user of the detectors being ready. 2. You will install the software/firmware and perform testing to ensure the software update was effective. 3.You will provide a customer notification letter that includes a statement that you will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. 2.Philips Healthcare will implement this CAP by November 10, 2017. If you need any further information or support concerning this Issue, please contact your local Philips representative: Technical Support Line: 1-800-722-9377. We apologize for any inconvenience this may cause and trust that this information is adequately addressing any concerns you may have.
Quantity
US - 132
Distribution
USA(nationwide) Distribution to the states of : AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, ME, MI, MN, MO, NC, ND, NE, NH, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA and WI
Lot, serial or model codes
DigitalDiagnost Upgrade Release 3.1.x to 3.1.3
510(k) numbers
["K141736"]
PMA numbers
not on file
Product code
MQB
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register