FDA Medical Device Recalls (openFDA)
Greiner Bio-One North America, Inc.: 3mL 9NC Coagulation Sodium Citrate 3.2%, Item #454334, Greiner Bio-One North America Inc., 4238 Capital Drive, Monroe, NC 28110. Intended Use: Venous blood collection tubes.
- Recall number
- Z-2154-2012
- Recalling firm
- Greiner Bio-One North America, Inc.
- Product
- 3mL 9NC Coagulation Sodium Citrate 3.2%, Item #454334, Greiner Bio-One North America Inc., 4238 Capital Drive, Monroe, NC 28110. Intended Use: Venous blood collection tubes.
- Status
- Terminated
- Initiated
- 2012-07-02
- Posted
- 2012-08-07
- Terminated
- 2013-06-11
- Reason
- Part of the lot received a lower concentration of citrate solution which may cause falsely lower values for common coagulation tests.
- Root cause
- Nonconforming Material/Component
- Action
- Distributors were notified by letter on 07/02/2012 and customers were sent a letter dated 07/05/2012. The letters identified the affected product and the reason for the recall. Distributors were asked to stop distributing the affected product and to complete the attached Product Disposition form. The form along with the affected product was to be returned to GBO. The firm will provide replacement stock or issue a credit once product return and receipt of the Product Disposition form has been confirmed. Customers were also asked to complete the Product Disposition Site Confirmation form and fax it back to Greiner Bio-One North America, Inc. at the number provided. The letter states that the firm will replace any affected product once the form is received. Questions should be directed to 800-515-8112.
- Quantity
- 838.9 cases (1,006,700 pieces)
- Distribution
- Nationwide Distribution
- Lot, serial or model codes
- Lot #B041206
- 510(k) numbers
- ["K971221"]
- PMA numbers
- not on file
- Product code
- GIM
- Firm FEI
- 3001451379
- Firm city
- Monroe
- Firm state
- NC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28