Skip to the content

FDA Medical Device Recalls (openFDA)

Acumed LLC: Acu-Loc Distal Radius Plate, Standard, Left. Product labeled in part, " Acu-Loc Distal Radius Plate, Standard, Left... PL-DR50L W48003... Manufactured or Distributed by: Acumed LLC Hillsboro, OR ".

Recall number
Z-2153-2008
Recalling firm
Acumed LLC
Product
Acu-Loc Distal Radius Plate, Standard, Left. Product labeled in part, " Acu-Loc Distal Radius Plate, Standard, Left... PL-DR50L W48003... Manufactured or Distributed by: Acumed LLC Hillsboro, OR ".
Status
Terminated
Initiated
2007-06-29
Posted
2008-08-28
Terminated
2010-02-03
Reason
Plates are incorrectly laser marked. The PL-DR50L plates are incorrectly laser marked with PL-DR70L.
Root cause
Process control
Action
The firm issued an Urgent Notification: Product Removal letter on 6-29-2007. The notification was reportedly sent by either FAX or e-mail. The notification informed consignees to remove the plates from their inventory and return them to Acumed. The notification did not include any instructions for a sub-recall. On 6/13/08, the firm sent a second Urgent Notification: Product Removal letter. This letter describes the product as PL-DR50L - Lot Code - W48003, explains the problem and asks customers to return product to Acumed.
Quantity
51 pieces
Distribution
Worldwide Distribution --- including USA and countries of South Africa, Italy, Finland, and Brazil.
Lot, serial or model codes
Lot: W4800; Expiration date: 2012-02
510(k) numbers
["K012655"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1000125930
Firm city
Hillsboro
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register