Biomet Spine, LLC: Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only, REF 14-583020, Biomet Spine and Bone Healing Technologies. Spinal fixation component.

Recall number
Z-2152-2015
Recalling firm
Biomet Spine, LLC
Product
Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only, REF 14-583020, Biomet Spine and Bone Healing Technologies. Spinal fixation component.
Status
Terminated
Initiated
2015-03-12
Posted
2015-07-20
Terminated
2015-12-21
Reason
Multiaxial screws may not meet internal requirements related to fatigue strength. Compression testing result and surface treatment are out of specification.
Root cause
Nonconforming Material/Component
Action
Biomet Spine issued an Urgent Medical Device Recall Notice/Response Form dated 3/12/2015 to their affected customers.
Quantity
4,765 units
Distribution
Distribution US Nationwide (AL, CA, CO, FL, GA, LA, MA, MD, MI, MO, NY, OH, PA, TX, WA, WI) and The Netherlands.
Lot, serial or model codes
Part Numbers: 14-582020, 14-582022, 14-582025, 14-582030, 14-582035, 14-582040, 14-582120, 14-582122, 14-582125, 14-582127, 14-582130, 14-582135, 14-582140, 14-582220, 14-582222, 14-582225, 14-582230, 14-582235, 14-582240, 14-582245, 14-582250, 14-592320, 14-582325, 14-582330, 14-582335, 14-582340, 14-582345, 14-582350, 14-582530, 14-582535, 14-582540, 14-582545, 14-582550, 14-582565, 14-582570, 14-582575, 14-583020, 14-583025, 14-583030, 14-583035, 14-583040, 14-583120, 14-583125, 14-583130, 14-583135, 14-583140, 14-583220, 14-583225, 14-583230, 14-583235, 14-583240, 14-583245, 14-583250, 14-583320, 14-583325, 14-583330, 14-583335, 14-583340, 14-583345 and 14-583350. All lots.
510(k) numbers
["K133746"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3010928172
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register