Skip to the content

FDA Medical Device Recalls (openFDA)

Impact Instrumentation, Inc.: Model 731 EMV+ Uni-Vent, Impact Instrumentation Inc., 27 Fairfield Place, West Caldwell NJ 07006. Indicated for use in the management of acute or chronic respiratory failure or during resuscitation by providing continuous positive -pressure ventilation.

Recall number
Z-2152-2012
Recalling firm
Impact Instrumentation, Inc.
Product
Model 731 EMV+ Uni-Vent, Impact Instrumentation Inc., 27 Fairfield Place, West Caldwell NJ 07006. Indicated for use in the management of acute or chronic respiratory failure or during resuscitation by providing continuous positive -pressure ventilation.
Status
Terminated
Initiated
2012-03-19
Posted
2012-08-07
Terminated
2012-08-15
Reason
Ten units of Model EMV+ Devices were shipped out in error before the verification and validation was completed for a new integrated circuit (IC) on the CPU PCB.
Root cause
Process control
Action
Impact Instrumentation Inc. notified their sole customer via email on 3/15/12 and then emailed an "Urgent Device Product Recall" letter and tracking form dated 3/19/2012. The letter described the reason for the recall and provided recommended actions and the contact number for Customer Service.
Quantity
Ten Units
Distribution
US Distribution to the state of Illinois.
Lot, serial or model codes
Serial numbers: VCDB120315, VCDB120316, VCDB120299, VCDB120300, VCDB120313, VCDB120307, VCDB120312, VCDB120303, VCDB120310, VCDB120297.
510(k) numbers
["K103318"]
PMA numbers
not on file
Product code
CBK
Firm FEI
1220908
Firm city
West Caldwell
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register