Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Knee Products: 189260 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 71 MM 189720 Vanguard Knee System, CR Mono Lock Tibial Bearing, 10 MM X 83 MM 189320 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 83 MM 189700 Vanguard Knee System, CR Mono Lock Tibial Bearing, 10 MM X 79 MM 189704 Vanguard Knee System, CR Mono Lock Tibial Bearing, 14 MM X 79 MM Product Usage: Knee prosthesis

Recall number
Z-2151-2020
Recalling firm
Biomet, Inc.
Product
Knee Products: 189260 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 71 MM 189720 Vanguard Knee System, CR Mono Lock Tibial Bearing, 10 MM X 83 MM 189320 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 83 MM 189700 Vanguard Knee System, CR Mono Lock Tibial Bearing, 10 MM X 79 MM 189704 Vanguard Knee System, CR Mono Lock Tibial Bearing, 14 MM X 79 MM Product Usage: Knee prosthesis
Status
Terminated
Initiated
2020-04-17
Posted
not on file
Terminated
2021-06-25
Reason
Potential presence of elevated endotoxin levels that exceed the specification limit
Root cause
Process control
Action
Zimmer Biomet issued Urgent Medical Device recall letter dated April 17, 2002 to distributors with direct sales and/or field inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory, as well as hospitals that have consumed products will be notified via FedEx. Surgeons that have consumed products will be notified via FedEx. Action to take: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form and send to CorporateQuality.PostMarket@zimmerbiomet.com within three (3) days. This form must be returned even if you do not have affected products available to return in your territory. b. For international returns, request an IRA by emailing zimmerbiometintlirarequests@zimmerbiomet.com c. For each return, send a copy of Attachment 1 to CorporateQuality.PostMarket@zimmerbiomet.com. CF04107 Rev. 4, Eff. Date: 16 Jul 2019 Ref. CP04102 Field Action Activities d. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. e. Mark RECALL on the outside of the returned cartons. 4. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or i
Quantity
11 units
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, AHMADABAD, GUJARAT, BANGKOK, BELROSE NSW, BUENOS AIRES, CHILE, COSTA RICA, HIRATSUKA CITY, HUECHURABA SANTIAGO, INDIA, JAPAN, MEXICO DF, NETHERLANDS, PICHINCHA, EC, SAN JOSE, SINGAPORE, TAIWAN, THAILAND, YONGSAN-GU SEOUL.
Lot, serial or model codes
Lot Numbers: UDI Number 630770 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 71 MM (01)00880304448490(17)211228(10)630770 646170 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 71 MM (01)00880304448490(17)211227(10)646170 530980 Vanguard Knee System, CR Mono Lock Tibial Bearing, 10 MM X 83 MM (01)00880304556140(17)211228(10)530980 570570 Vanguard Knee System, CR-L Mono Lock Tibial Bearing, 10 MM X 83 MM (01)00880304448605(17)220106(10)570570 477430 Vanguard Knee System, CR Mono Lock Tibial Bearing, 10 MM X 79 MM (01)00880304493131(17)220115(10)477430 505490 Vanguard Knee System, CR Mono Lock Tibial Bearing, 14 MM X 79 MM (01)00880304556119(17)220115(10)505490
510(k) numbers
["K060525"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register