Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Vascular: Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.

Recall number
Z-2151-2017
Recalling firm
Abbott Vascular
Product
Absorb Bioresorbable Vascular Scaffold (BVS) system (Australia). The Absorb BVS is a temporary scaffold indicated for improving coronary luminal diameter that will eventually resorb and potentially facilitate normalization of vessel function in patients with ischemic heart disease due to de novo native coronary artery lesions. The treated lesion length should be less than the nominal scaffolding length (8 mm, 12 mm, 18 mm, 23 mm, 28 mm) with reference vessel diameters e 2.0 mm and d 3.8 mm.
Status
Terminated
Initiated
2017-04-26
Posted
2017-06-01
Terminated
2017-11-16
Reason
Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent.
Root cause
Other
Action
Abbott Vascular sent an Urgent Medical Device Recall and Hazard Alert letter dated April 26, 2017, to affected customers to inform them that the action is in response to recent concerns over data from some studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated with the Xience metallic drug eluting stent. Customers were informed of the actions to be taken. Customers were instructed to stop using these devices immediately, advise BVS patients to follow the recommendations for DAPT prescribed by their health care provider. Advise patients experiencing new cardiac symptoms to seek clinical care, review their inventory and complete the attached Facsimile Reply Form and display and share this notification with other relevant personnel within their organization. Customers with questions are instructed to contact their local Abbott Vascular Representative or Customer Service department.
Quantity
590 units
Distribution
Australia
Lot, serial or model codes
Australia: Part No. Description 1012462-08 2.5 x 08 mm Absorb BVS 1012462-12 2.5 x 12 mm Absorb BVS 1012462-18 2.5 x 18 mm Absorb BVS 1012462-23 2.5 x 23 mm Absorb BVS 1012462-28 2.5 x 28 mm Absorb BVS 1012463-08 3.0 x 08 mm Absorb BVS 1012463-12 3.0 x 12 mm Absorb BVS 1012463-18 3.0 x 18 mm Absorb BVS 1012463-23 3.0 x 23 mm Absorb BVS 1012463-28 3.0 x 28 mm Absorb BVS 1012464-12 3.5 x 12 mm Absorb BVS 1012464-18 3.5 x 18 mm Absorb BVS 1012464-23 3.5 x 23 mm Absorb BVS 1012464-28 3.5 x 28 mm Absorb BVS
510(k) numbers
not on file
PMA numbers
["P150023"]
Product code
PNY
Firm FEI
2024168
Firm city
Temecula
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register