Skip to the content

FDA Medical Device Recalls (openFDA)

Linvatec Corp. dba ConMed Linvatec: PopLok Punch labeled in part: "PKL-35M***3.5mm***QTY 1***PopLok Punch***ConMed Linvatec Biomaterials, Ltd. Hermiankatu 6-8L 33720 Tampere, FINLAND***Rx Only***ConMed Linvatec 11311 Concept Blvd Largo, FL 33773-4908 USA***Made in USA***" Intended Use: Bone Punch.

Recall number
Z-2151-2012
Recalling firm
Linvatec Corp. dba ConMed Linvatec
Product
PopLok Punch labeled in part: "PKL-35M***3.5mm***QTY 1***PopLok Punch***ConMed Linvatec Biomaterials, Ltd. Hermiankatu 6-8L 33720 Tampere, FINLAND***Rx Only***ConMed Linvatec 11311 Concept Blvd Largo, FL 33773-4908 USA***Made in USA***" Intended Use: Bone Punch.
Status
Terminated
Initiated
2012-07-06
Posted
2012-08-06
Terminated
2013-11-21
Reason
ConMed Linvatec recalled one lot of their PopLok bone Punch 3.5MM because the metal shaft diameter is 4.5mm when it should have been 3.5mm.
Root cause
Process change control
Action
The recall is being conducted to the end user level. ConMed Linvatec Urgent Medical Device Recall Notifications were sent by FedEx next day delivery or equivalent method to notify all direct consignees. The notifications identified the affected product along with the reason for the recall and health risk. All accounts were asked to immediately check their facility for the affected product, stop using any of the product with specified lot number, and to segregate the product. The letter also asked customers to return all affected product to ConMed Linvatec and respond using a Reply Form, even if there is no affected product available. If customers have any additional questions, they should contact the Customer Service department at 800-237-0169.
Quantity
2
Distribution
Worldwide Distribution -- US and Korea.
Lot, serial or model codes
Lot 318455
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1017294
Firm city
Largo
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register