Skip to the content

FDA Medical Device Recalls (openFDA)

GE Medical Systems, LLC: Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6

Recall number
Z-2150-2026
Recalling firm
GE Medical Systems, LLC
Product
Revolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
Status
Open, Classified
Initiated
2026-03-26
Posted
2026-05-08
Terminated
not on file
Reason
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
Root cause
Software design
Action
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to recall.23015@gehealthcare.com Product Correction GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare
Quantity
38 units
Distribution
Worldwide - US Nationwide distribution.
Lot, serial or model codes
UDI-DI: 00840682118552; Serial Numbers:REVVX2300007CN 443327CN2 REVVX1600034CN REVVX2400004CN REVVX2200047CN 441716CN8 00000440713CN6 REVVX1800043CN 441839CN8 REVVX1700067CN REVVX1800007CN REVVX1800105CN REVVX1800048CN REVVX1900123CN REVVX1800046CN REVVX2000005CN REVVX2200009CN REVVX1600013CN REVVX1700101CN REVVX2200025CN REVVX2100028CN REVVX1900069CN REVVX2100020CN REVVX2100014CN REVVX2100050CN REVVX1600052CN 438902CN9 REVVX1900107CN 438369CN1 REVVX1900018CN REVVX1700071CN REVVX1700082CN 442792CN8 REVVX2200007CN REVVX1700104CN REVVX2000056CN REVVX1900086CN REVVX2200039CN
510(k) numbers
["K133705","K163213","K191777","K213715"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register