Varian Medical Systems, Inc.: Optical Surface Monitoring System{Align RT Plus} This is used on Varian's TrueBeam EDGE; Radiology: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to Withhold the beam automatically during radiation delivery.

Recall number
Z-2150-2016
Recalling firm
Varian Medical Systems, Inc.
Product
Optical Surface Monitoring System{Align RT Plus} This is used on Varian's TrueBeam EDGE; Radiology: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to Withhold the beam automatically during radiation delivery.
Status
Terminated
Initiated
2016-06-23
Posted
2016-07-11
Terminated
2018-08-24
Reason
Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different couch rotation from the plan couch rotation while performing patient alignment with Optical Surface Monitoring System [OSMS].
Root cause
Software design
Action
The firm, Visionrt, sent an "URGENT MEDICAL DEVICE CORRECTION NOTIFICATION-URGENT FIELD SAFETY NOTIFICATION" letter dated June 24, 2016 to all affected customers. Letters describe the product, problem and actions to be taken. The customers were instructed as follows: DO NOT USE the "Move Couch" function if the couch rotation value set in the AlignRT device DOES NOT match the planned couch rotation for the treatment field; retain a copy of this document along with your most current product labeling; pass the letter to all those who need to be aware of it within their organization; always confirm that the patient is correctly aligned prior to commencing treatment; is not required to return their devices; promptly inform Vision RT if they believe that patient harm occurred, and complete Appendix 1- Customer Acknowledgment Response and return it via email to corrections@visionrt.com. Should you have any queries on this letter, please do not hesitate to contact Vision RT by telephone on +44 20 83464300 (866-778-2379 from the US) or as per http://www.visionrt.com/contact/details.
Quantity
21 AlignRT Plus in US, 1 International.
Distribution
Worldwide Distribution: US (nationwide) and countries of: Austria, Azerbaijan, Brunei Darussalam, Bulgaria, Denmark, Georgia, Germany, India, Italy, Malaysia, Morocco, Norway, Portugal, Switzerland, and United Kingdom.
Lot, serial or model codes
TrueBeam version 2.0/2.5 with any version OSMS.  AlignRT Serial Numbers: 247-0200 247-0209 248-0013 248-0057 248-0070 249-0057 249-0066 249-0163 249-0164 249-0195 249-0226 249-0274 249-0050 249-0286 249-0277 248-0044 249-0227 247-0208 249-0047 247-0218 248-0050 248-0043 248-0048 249-0293 249-0155.
510(k) numbers
["K123371"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register