Siemens Medical Solutions USA, Inc.: Siemens brand ACUSON S2000, Model 10434730, ABVS ultrasound systems; Affected sub-components Numbers: (1) 10434968, Boom Welding (for S/Ns 250157, 250162, 250163, 250164, 250166, 250167 and 250169); (2) 10434969, Arm Welding (for S/N's 250171, 250173, 250177); Product is manufactured and distributed by Siemens Medical Solutions USA, Inc., Concord, CA The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging the breast allows volumetric acquisition of the breast image. It provides the operator with the option of using a controlled user independent automated scan motion. The resulting 3D display information enables comprehensive image review by the medical professional. In particular, the coronal plane is scanned, which is not easily accessible with hand-held ultrasound transducers.

Recall number
Z-2150-2011
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens brand ACUSON S2000, Model 10434730, ABVS ultrasound systems; Affected sub-components Numbers: (1) 10434968, Boom Welding (for S/Ns 250157, 250162, 250163, 250164, 250166, 250167 and 250169); (2) 10434969, Arm Welding (for S/N's 250171, 250173, 250177); Product is manufactured and distributed by Siemens Medical Solutions USA, Inc., Concord, CA The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging the breast allows volumetric acquisition of the breast image. It provides the operator with the option of using a controlled user independent automated scan motion. The resulting 3D display information enables comprehensive image review by the medical professional. In particular, the coronal plane is scanned, which is not easily accessible with hand-held ultrasound transducers.
Status
Terminated
Initiated
2011-03-28
Posted
2011-05-06
Terminated
2011-05-06
Reason
The weld connecting the ABVS arm to the system base has the potential to fail, causing the arm to fall off of the unit.
Root cause
Process change control
Action
Siemens Medical Solutions USA, Inc. initiated the recall via phone calls and e-mail on March 28, 2011, to all affected customers, informing them of the affected product and providing instructions on the recall. Three units S/N 250171, 250173, and 250177 were to be removed and shipped back to Siemens. Siemens Service personnel were to be dispatched to the customer sites to perform validated stress test on units S/N 250157, 250162, 250163, 250164, 250166, 250167, and 250169 at customer sites. For questions regarding this recall call 650-694-5993.
Quantity
10
Distribution
Worldwide Distribution - USA, Portugal, Germany, Russia, Estonia, Brazil, France, and Saudi Arabia.
Lot, serial or model codes
Serial Numbers Affected: 250171, 250173, 250177, 250157, 250162, 250163, 250164, 250166, 250167 and 250169
510(k) numbers
["K081148"]
PMA numbers
not on file
Product code
IYN
Firm FEI
2936884
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register