Maquet Cardiovascular Us Sales, Llc: Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794

Recall number
Z-2149-2019
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Product
Maquet Cardiopulmonary (MCP)/Getinge HLS Set Advanced 7.0. Model Number BEQ-HLS-7050 USA, Part Number 701052794
Status
Terminated
Initiated
2019-07-11
Posted
not on file
Terminated
2020-12-11
Reason
The sets are configured with quick connectors that have been assembled in reverse on the arterial and venous lines, resulting in blue (venous) to red (arterial) and red (arterial) to blue (venous) connections. Due to this incorrect assembly, the sets cannot be primed from the reservoir prior to use, and the device may not function as intended.
Root cause
Under Investigation by firm
Action
Urgent Medical Device Recall - Removal notification letters dated 7/11/19 were sent to customers.
Quantity
54
Distribution
The products were distributed to the following US states: AR, CA, FL, GA, IA, IL, IN, KY, NC, NE, NM, NV, NY, OH, OR, TN, and WI.
Lot, serial or model codes
Lot 70131093; UDI 04037691741543
510(k) numbers
["K112360"]
PMA numbers
not on file
Product code
DTZ
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register