EKOS Corporation: The EKOS EkoSonic Control Unit is intended exclusively for use with the EkoSonic Endovascular Device and the EkoSonic MACH4 Endovascular Device. The EkoSonic Endovascular System consists of three main components: -a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and -removable MicroSonic Device (MSD), and -a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, nonsterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.

Recall number
Z-2148-2015
Recalling firm
EKOS Corporation
Product
The EKOS EkoSonic Control Unit is intended exclusively for use with the EkoSonic Endovascular Device and the EkoSonic MACH4 Endovascular Device. The EkoSonic Endovascular System consists of three main components: -a single use EkoSonic Endovascular Device which is comprised of an Intelligent Drug Delivery Catheter (IDDC) and -removable MicroSonic Device (MSD), and -a reusable EKOS EkoSonic Control System. During use, fluids are delivered through the side-holes of the IDDC. The MSD is placed through the central lumen of the IDDC to deliver ultrasound energy along the length of fluid delivery, dispersing the fluid into the intravascular treatment site. The reusable EKOS EkoSonic Control Unit provides power to the system and the user interface for operator control. A reusable, nonsterile Connector Interface Cable connects the EKOS EkoSonic Control Unit to the EkoSonic Endovascular Device.
Status
Terminated
Initiated
2015-06-24
Posted
2015-07-20
Terminated
2015-08-25
Reason
The Connector Interface Cable (CIC) was not recognized by the EkoSonic PT-3B control unit.
Root cause
Vendor change control
Action
Ekos sent an "Urgent Voluntary Medical Device and Replacement" letter dated June 24, 2015, to US customers. The letter identified the product the problem and the action needed to be taken by the customer. The replacement CIC and recall letter were sent by FEDERAL EXPRESS directly to the person responsible for maintaining the EkoSonic System. Tracking numbers and delivery notifications will be retained for each shipment. Letters to Risk Management and Purchasing were sent by REGISTER MAIL. The return cards will be retained for each letter. Customers with questions can contact EKOS Customer Service at 1-888-400-3567.
Quantity
46 units
Distribution
Nationwide Distribution
Lot, serial or model codes
CIC Serial Numbers:  19598-095, 19598-097, 19598-099, 19598-100, 19598-102, 19598-103,  19598-106, 19598-107, 19598-108, 19598-110, 19598-111, 19598-112,  19598-113, 19598-114, 19598-115, 19598-116, 19598-117, 19598-118,  19598-119, 19598-120, 19598-122, 19598-125, 19598-126, 19598-127,  19598-128, 19598-130, 19598-131, 19598-132, 19598-133, 19598-135,  19598-136, 19598-139, 19598-140, 19598-141, 19598-143, 19598-144,  19598-145, 19598-146, 19598-147, 19598-151, 19598-152, 19598-153,  19598-154, 19598-156, 19598-157, 19598-158.
510(k) numbers
["K081467","K111705","K140151"]
PMA numbers
not on file
Product code
KRA
Firm FEI
3001627457
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register