Skip to the content

FDA Medical Device Recalls (openFDA)

Genzyme Corporation: Equal Diagnostics Lipase Color Reagent Kit Part Number: 905-B For In Vitro Diagnostic Use Containing the Lipase Color Reagent, 30 ml, Part Number 80-2005-01.

Recall number
Z-2148-2008
Recalling firm
Genzyme Corporation
Product
Equal Diagnostics Lipase Color Reagent Kit Part Number: 905-B For In Vitro Diagnostic Use Containing the Lipase Color Reagent, 30 ml, Part Number 80-2005-01.
Status
Terminated
Initiated
2008-04-14
Posted
2008-08-22
Terminated
2008-09-16
Reason
Calibration and Quality Control outside of established range resulting from an increase in initial OD readings with the reagent contained within the kits.
Root cause
Nonconforming Material/Component
Action
Genzyme notified accounts by Fed'X on 4/14/08 via an Imprtant Field Correction letter, requesting users to discard kits and follow the "User Update" instructions previously provided to minimize calibration error. Customers are to complere and sign the Confirmation of Notification form and fax it back to the firm regardless if they have any affected product or not. Also, requires that the customer confirms that the REagent OD limits for First and Last H are set at 0.15 for Lipase on the Olympus AU400/400e, AU600, AU640/640e and AU2700/AU5400 Systems.
Quantity
3,443
Distribution
Nationwide Distribution
Lot, serial or model codes
Reagent lot Number 491686, Lot #: 7EPM30 exp 01/15/09
510(k) numbers
["K897150"]
PMA numbers
not on file
Product code
CHI
Firm FEI
1221955
Firm city
Cambridge
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register