FDA Medical Device Recalls (openFDA)

GE Healthcare: GE Stereotaxy Positioner, model 2405544-3, for use with Senographe DS Full Field Mammography system, models 2383168, 2383168-2, 2383168-3, 2383168-3-1, 2383168-4-1. The expected use of the Senographe DS Stereotaxy is an optional accessory for the Senographe system for mammography examinations.

Recall number
Z-2147-2008
Recalling firm
GE Healthcare
Product
GE Stereotaxy Positioner, model 2405544-3, for use with Senographe DS Full Field Mammography system, models 2383168, 2383168-2, 2383168-3, 2383168-3-1, 2383168-4-1. The expected use of the Senographe DS Stereotaxy is an optional accessory for the Senographe system for mammography examinations.
Status
Terminated
Initiated
2008-07-01
Posted
2008-09-26
Terminated
2011-12-11
Reason
GE Healthcare has recently become aware of x-ray emission beyond the edge of the detector primary barrier. This issue occurs when an exam is performed in a specific angulated view associated with use of the Stereotactic Positioner of your Senographe DS Acquisition system and could impact patient safety. If the Stereotactic Positioner is removed when the examination arm is at 33 degrees and, th
Root cause
Radiation Control for Health and Safety Act
Action
A Product Safety Notification letter, dated March 3, 2008, was sent to affected customers. The letter provided information about the safety issue, identified affected product and stated that GE Healthcare will provide a software upgrade. Contact GE Healthcare at 1-262-544-3894 for assistance.
Quantity
12
Distribution
AL, CA, CO, FL, IL, KS, MA, ME, MN,MO, NC, NJ, NM, NY OK, SC, TN, TX, and WI.
Lot, serial or model codes
model 2405544-3, Serial numbers: 00000509762BU1, 00000509764BU7, 00000543767BU8, 00000546136BU3, 00000546140BU5, 00000546772BU5, 00000546793BU1, 00000546794BU9, 00000546799BU8, 00000550251BU3, 00000550265BU3, and 00000550266BU1.
510(k) numbers
["K941191","K990066"]
PMA numbers
not on file
Product code
IZH
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register