Ortho-Clinical Diagnostics: enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; IVD

Recall number
Z-2143-2016
Recalling firm
Ortho-Clinical Diagnostics
Product
enGen Laboratory Automation System, Product Code engen, Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; IVD
Status
Terminated
Initiated
2016-05-16
Posted
2016-07-08
Terminated
2018-06-08
Reason
If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends the incorrect text to the LIS. The IFU states the interpretation term for these numerical result values is Retest?. The difference in terminology exists for US Markets only.
Root cause
Software design
Action
Ortho Clinical Diagnostics sent an "IMPORTANT PRODUCT CORRECTION NOTIFICATION" Customer Letter dated May 16, 2016, to all affected customers in the US with enGen Systems via FedEx or US Priority Mail (for Post Office Box addresses) or by e-mail to subscribers of ORTHO PLUS e-Communications to advise them of the issue and the required actions . Required Actions: 1. Use the numerical test results to assess actions to be taken for these assays, per the IFU. 2. Post this notification by your enGen System or with your user documentation. 3. Complete and return the Confirmation of Receipt form by 25 May 2016. with the Instructions for Use (IFU) directions). If you have any questions regarding this notification, please call the Ortho Technical Solutions Center at any time at 1-800-421-3311.
Quantity
US: 30 units
Distribution
Nationwide
Lot, serial or model codes
Software Versions enGen Select 3.2 and Above and enGen Custom 2.0 and Above; Serial Numbers/J Numbers: JIM341866, JIM341820, JIM341679, JIM341352, JIM343421, JIM341174, JIM336745, JIM340667, JIM338283,  JIM-341592, JIM336957, JIM341996, JIM342334, JIM343694, JIM336855,  JIM334236, JIM341415, JIM341392, JIM336481, JIM340934, JIM336167,  JIM342249, JIM338290, JIM341750, JIM341426, JIM334986, JIM334527,  JIM340024, JIM341021, JIM340658
510(k) numbers
["K063144"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register