Fresenius Medical Care Holdings, Inc.: US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Recall number
Z-2141-2024
Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1
Status
Open, Classified
Initiated
2024-05-16
Posted
2024-06-13
Terminated
not on file
Reason
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
Root cause
Process control
Action
Fresenius Medical issued Urgent Medical Device Correction letter on 5-16-2024. Letter states reason for recall, health risk and action to take: A Fresenius Medical Care Water Systems Service Specialist will be contacting you to schedule an evaluation of your Water Treatment system and provide feedback on your system. For questions or concerns, please contact Fresenius Medical Care Technical Services at 800-227-2572, OPTION 4. " Please provide this letter to your Biomedical Technician " Report any complaints or adverse events to product.complaints@fmc-na.com or FDA MedWatchathttps://www.fda.gov/safety/medical-product-safety-information/medwatch-forms-fda-safety-reporting. " Provide this Customer Notice to all those who need to be aware within your organization. " Complete and return the attached reply form.
Quantity
4 units
Distribution
US Nationwide distribution in the states of AL, IN, AK, WI.
Lot, serial or model codes
UDI-DI: 04251285200357 Serial Numbers: 2AAV0003 2AAV0029 2AAV0037 3AAV0087
510(k) numbers
["K143617"]
PMA numbers
not on file
Product code
FIP
Firm FEI
3001451489
Firm city
Waltham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register