Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific NexStent Monorail, NexStent Carotid Stent and Monorail Delivery System. Material Number (UPN) M001553030, Catalog Number 55-303. [OUS Material Number (UPN) M001553000, Catalog Number 55-300]

Recall number
Z-2141-2008
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific NexStent Monorail, NexStent Carotid Stent and Monorail Delivery System. Material Number (UPN) M001553030, Catalog Number 55-303. [OUS Material Number (UPN) M001553000, Catalog Number 55-300]
Status
Terminated
Initiated
2008-06-06
Posted
2008-08-13
Terminated
2011-02-16
Reason
Tip detachment: Boston Scientific has received complaints involving a detachment of the tip from the NexStent Carotid Stent delivery system. Potential health hazards resulting from this type of failure include increased procedure time, vessel wall injury, stroke and/or emergency surgery to remove the detached tip. Since the potential for the tip to detach occurs during delivery of the sten
Root cause
Component design/selection
Action
Boston Scientific sent an urgent Voluntary Medical Device Recall letter, dated 06/06/08, to affected customers. The letter described the issue, identified affected product, requested to cease using or distributing the product and provided direction on how to return the recalled devices to Boston Scientific, Quincy, MA. A Reply Verification Tracking Form was asked to be returned to Boston Scientific. Since the issue is with the delivery system, patients already implanted with the stent are not affected.
Quantity
1946 units (1570 U.S., 376 OUS)
Distribution
Worldwide Distribution: USA, Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Hungary, Italy, Kuwait, Luxembourg, Netherlands, Norway, Panama, Poland, Portugal, Romania, Spain, Switzerland, and Ukraine.
Lot, serial or model codes
US Lot/Batch #: C71403, C71404, C71405, C71406, C71501, C71502, C71601, C71602, C72302, C72303, C72401, C72402, C72403, C72404, C72701, C72801, C72802, C72803, C72901, C72902, C72903, C72904, C73002, C73003, C73103, C73104, C73105, C73201, C73202, C73203, C73204, C73205, C73301, C73302, C73401, C73402, C73601, C73602, C73801, C74301, C74302, C74401, C74502, C74601, C74701, C80201, C80202, C80203, C80301, C80401, C80402, C80403, C80501, C80801, C80802, C81001, C81002, C81102, C81201, C81202, and C81203.   OUS Lot/Batch #: C72501, C72502, C72601, C72602, C72905, C73001, C73004, C73101, C73102, C73901, C73902, C73903, C74001, C74002, C74003, and C74101
510(k) numbers
not on file
PMA numbers
["P050025"]
Product code
NIM
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register