FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: Boston Scientific NexStent Monorail, NexStent Carotid Stent and Monorail Delivery System. Material Number (UPN) M001553030, Catalog Number 55-303. [OUS Material Number (UPN) M001553000, Catalog Number 55-300]
- Recall number
- Z-2141-2008
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific NexStent Monorail, NexStent Carotid Stent and Monorail Delivery System. Material Number (UPN) M001553030, Catalog Number 55-303. [OUS Material Number (UPN) M001553000, Catalog Number 55-300]
- Status
- Terminated
- Initiated
- 2008-06-06
- Posted
- 2008-08-13
- Terminated
- 2011-02-16
- Reason
- Tip detachment: Boston Scientific has received complaints involving a detachment of the tip from the NexStent Carotid Stent delivery system. Potential health hazards resulting from this type of failure include increased procedure time, vessel wall injury, stroke and/or emergency surgery to remove the detached tip. Since the potential for the tip to detach occurs during delivery of the sten
- Root cause
- Component design/selection
- Action
- Boston Scientific sent an urgent Voluntary Medical Device Recall letter, dated 06/06/08, to affected customers. The letter described the issue, identified affected product, requested to cease using or distributing the product and provided direction on how to return the recalled devices to Boston Scientific, Quincy, MA. A Reply Verification Tracking Form was asked to be returned to Boston Scientific. Since the issue is with the delivery system, patients already implanted with the stent are not affected.
- Quantity
- 1946 units (1570 U.S., 376 OUS)
- Distribution
- Worldwide Distribution: USA, Austria, Belgium, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Hungary, Italy, Kuwait, Luxembourg, Netherlands, Norway, Panama, Poland, Portugal, Romania, Spain, Switzerland, and Ukraine.
- Lot, serial or model codes
- US Lot/Batch #: C71403, C71404, C71405, C71406, C71501, C71502, C71601, C71602, C72302, C72303, C72401, C72402, C72403, C72404, C72701, C72801, C72802, C72803, C72901, C72902, C72903, C72904, C73002, C73003, C73103, C73104, C73105, C73201, C73202, C73203, C73204, C73205, C73301, C73302, C73401, C73402, C73601, C73602, C73801, C74301, C74302, C74401, C74502, C74601, C74701, C80201, C80202, C80203, C80301, C80401, C80402, C80403, C80501, C80801, C80802, C81001, C81002, C81102, C81201, C81202, and C81203. OUS Lot/Batch #: C72501, C72502, C72601, C72602, C72905, C73001, C73004, C73101, C73102, C73901, C73902, C73903, C74001, C74002, C74003, and C74101
- 510(k) numbers
- not on file
- PMA numbers
- ["P050025"]
- Product code
- NIM
- Firm FEI
- 3002095335
- Firm city
- Maple Grove
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as