Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Medical Systems USA: Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.

Recall number
Z-2140-2021
Recalling firm
Maquet Medical Systems USA
Product
Heater Cooler Unit (HCU 30) The device supplies temperature-controlled water for cardioplegia heat exchangers and for blood heat exchangers in extracorporeal circulation and for blankets with which patients can be warmed or cooled.
Status
Terminated
Initiated
2021-07-14
Posted
not on file
Terminated
2022-08-03
Reason
There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species.
Root cause
Device Design
Action
On July 14, 2021, Maquet Getinge Group issued Urgent Medical Device Recall notices to customers via FedEx 2 day Delivery with Signature Proof of Delivery. Customers were advised to take the following action: - Users should take the unit out of operation at the earliest opportunity and decommission the unit. - MCP has not developed a cleaning protocol that meets current concerns and expectations of the FDA. - If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. - Please acknowledge receipt of the Urgent Medical Device Recall by completing and returning the response form. Please either fax the completed form to (866) 5715830 or send via email to MCPHCU30.qrc@getinge.com.
Quantity
33
Distribution
Distribution to US states of AL, CA, FL, IL, KS, MI, MO, NC, NE & OH.
Lot, serial or model codes
HCU 30 Base Unit 100-120 V, Model Number 704629 I 70103.4371 HCU 30 Base Unit Canadian/US, Model Number 705501 I 70103.4653HCU 30 Base Unit 200-240 V, Model Number 939001 I 70102.8718 All serial numbers are affected.
510(k) numbers
["K031544"]
PMA numbers
not on file
Product code
DWC
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register