DePuy Orthopaedics, Inc.: ATTUNE CONV RP CR ARTICULATION SURFACE SZ1, 2, 9, and 10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.

Recall number
Z-2140-2015
Recalling firm
DePuy Orthopaedics, Inc.
Product
ATTUNE CONV RP CR ARTICULATION SURFACE SZ1, 2, 9, and 10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.
Status
Terminated
Initiated
2015-06-12
Posted
2015-07-20
Terminated
2015-11-12
Reason
Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
Root cause
Device Design
Action
URGENT INFORMATION: MEDICAL DEVICE CORRECTION was issued for all lots of the affected devices. The device correction is to inform US Distributors and Medical Professionals about the issue, how to prevent the issue, and any actions to take. US Distributors will provide Hospitals/User Facilities with the Medical Professional Notice and Acknowledgment Cards. US Distributors will also work with the sales consultants to ensure all consultants complete sales training regarding this issue.
Quantity
13964 units
Distribution
Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and Internationally to CANADA , AUSTRALIA, AUSTRIA, BELGIUM, ENGLAND/ UK, FRANCE, GERMANY, HONG KONG, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JAPAN, MALAYSIA, NETHERLANDS, NEW ZEALAND, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN & IBERIA, SWEDEN, SWITZERLAND, THAILAND, UAE, CHILE, COLOMBIA, MEXICO, PERU, and PUERTO RICO.
Lot, serial or model codes
Product Codes:254500981, 254500982, 254501989, and 254501990. GTIN: 10603295135586, 10603295135593, 10603295423492, and 10603295423959.
510(k) numbers
["K140881"]
PMA numbers
not on file
Product code
MBH
Firm FEI
1818910
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register