Skip to the content

FDA Medical Device Recalls (openFDA)

Acumed LLC: Osteo-Clage Cable System Package Assembly (01-0001-S). Product labeled in part, "2.0mm x 850 mm CoCr Cable / Sleeve... REF 01-0001-S LOT W59192... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR 97124-9432 ".

Recall number
Z-2140-2008
Recalling firm
Acumed LLC
Product
Osteo-Clage Cable System Package Assembly (01-0001-S). Product labeled in part, "2.0mm x 850 mm CoCr Cable / Sleeve... REF 01-0001-S LOT W59192... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR 97124-9432 ".
Status
Terminated
Initiated
2008-03-28
Posted
2008-08-22
Terminated
2010-02-03
Reason
Cables of incorrect diameter were used in product assembly, resulting in the sleeve not being able to bind to the cable after crimping.
Root cause
Employee error
Action
On 3/28/08, the firm sent an Urgent Notification letter to their consignees. The letter explained the issue and requested the customers return the product to Acumend. On 7/16/08, the firm sent a 2nd notification, via mail, to thier customers. The Urgent Notice: Device Recall 2nd Notification, advised customers of the recall, the hazard and requested customers to return existing product to Acumed.
Quantity
50 units
Distribution
Nationwide Distribution --- including states of AR, CA, PA, GA, KS, MO, and TN.
Lot, serial or model codes
Work order number/Lot: W59192 (W=work order, ###=sequential number); Expiration date: 2015-01 (Year/Month)
510(k) numbers
["K921480"]
PMA numbers
not on file
Product code
JDQ
Firm FEI
1000125930
Firm city
Hillsboro
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register