RAYSEARCH LABORATORIES AB: RayStation (Radiation Treatment Planning System) : RayStation 9A, 9B, 9B SP1, 10A, 10A SP1, 10B and 11A, in combination with RayCare Model Numbers/ UDI: RayStation 9.0 (9.0.0.113) 07350002010174; RayStation 9.1 (9.1.0.933) 07350002010266; RayStation 9.1 Service Pack 1 (9.2.0.483) 07350002010297; RayStation 10.0 (10.0.0.1154) 07350002010303; RayStation 10.0 Service Pack 1 (10.0.1.52) 07350002010365; RayStation 10.1 (10.1.0.613) 07350002010310; RayStation 11.0 (11.0.0.951) 07350002010389;

Recall number
Z-2138-2021
Recalling firm
RAYSEARCH LABORATORIES AB
Product
RayStation (Radiation Treatment Planning System) : RayStation 9A, 9B, 9B SP1, 10A, 10A SP1, 10B and 11A, in combination with RayCare Model Numbers/ UDI: RayStation 9.0 (9.0.0.113) 07350002010174; RayStation 9.1 (9.1.0.933) 07350002010266; RayStation 9.1 Service Pack 1 (9.2.0.483) 07350002010297; RayStation 10.0 (10.0.0.1154) 07350002010303; RayStation 10.0 Service Pack 1 (10.0.1.52) 07350002010365; RayStation 10.1 (10.1.0.613) 07350002010310; RayStation 11.0 (11.0.0.951) 07350002010389;
Status
Terminated
Initiated
2021-06-08
Posted
not on file
Terminated
2024-03-20
Reason
Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.
Root cause
Under Investigation by firm
Action
RaySearch issued Urgent Field Safety Notice, Medical Device Correction #80230 via email on 06/08/21. Letter states reason for recall, health risk and action to take: TO BE TAKEN BY THE USER " Do not create setup beams with the localization point as isocenter (option 2). " Always perform secondary verification of new treatment positions. RayTreat will notify the user when the treatment positions have been updated. " Always perform imaging before treating the patient. " Educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. SOLUTION This issue will be resolved in the next version of RayStation, scheduled for market release in June 2021 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact quality@raysearchlabs.com.
Quantity
5 units
Distribution
US distribution to TN; and Belgium
Lot, serial or model codes
Product name (build number) UDI-DI RayStation 9.0 (9.0.0.113) 07350002010174 RayStation 9.1 (9.1.0.933) 07350002010266 RayStation 9.1 Service Pack 1 (9.2.0.483) 07350002010297 RayStation 10.0 (10.0.0.1154) 07350002010303 RayStation 10.0 Service Pack 1 (10.0.1.52) 07350002010365 RayStation 10.1 (10.1.0.613) 07350002010310 RayStation 11.0 (11.0.0.951) 07350002010389
510(k) numbers
K200487, K200569
PMA numbers
not on file
Product code
MUJ
Firm FEI
3007774465
Firm city
Stockholm
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
3e6df46828c1
Size
288213574 bytes
Published
2026-10-06
Read
2026-10-07T03:20:25.099Z
Source file
Source file retained · verify
Refresh due
2026-10-14

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 3e6df46828c1, published 2026-10-06

Search this source