Synovis Micro Companies Alliance Inc: GEM 2753 Microvascular Anastomotic Coupler, 2.5 mm, REF GEM2753/I, P/N 5111-00250-010, Rx only, Irradiation Sterilized This is a microsurgical coupler used in the anastomosis of veins and arteries normally encountered in microsurgical vascular reconstructive procedures.
- Recall number
- Z-2138-2008
- Recalling firm
- Synovis Micro Companies Alliance Inc
- Product
- GEM 2753 Microvascular Anastomotic Coupler, 2.5 mm, REF GEM2753/I, P/N 5111-00250-010, Rx only, Irradiation Sterilized This is a microsurgical coupler used in the anastomosis of veins and arteries normally encountered in microsurgical vascular reconstructive procedures.
- Status
- Terminated
- Initiated
- 2008-05-14
- Posted
- 2008-08-02
- Terminated
- 2010-04-09
- Reason
- Coupler rings may slip out of the delivery toolcaused by jaws larger than specification.
- Root cause
- Device Design
- Action
- Urgent Recall Notification was sent via Fed Ex on May 15, 2008 to 432 hospitals. The letters were addressed to the attention of the Risk Manager. The recall notification instructed the hospital to immediately discontinue use, examine their inventory and, if found, to remove the product and quarantine it for return. A second notification was sent on July 21, 2008 to non-responding consignees.
- Quantity
- 3,812 units
- Distribution
- Worldwide Distribution --- including USA and countries of Canada, Japan, UAE, Korea, South Africa, Australia, Taiwan, Norway, Netherlands, Sweden, Greece, Norway, Turkey, Portugal, Taiwan, Denmark, Italy, Germany, Japan, Spain, Austria, and Great Britain.
- Lot, serial or model codes
- Lot numbers: 1808, 1894, 1907, 1923, 1953, 1954, 7139, 7211, 7354, 8998, 8999, 11591, 12364, 15244, 15280, 15281, 18824, 18912, 19294, 19413, 19954, 20263, 20689, 21345, 21515, 21560, 21977, 22556, and 23047
- 510(k) numbers
- ["K040163"]
- PMA numbers
- not on file
- Product code
- MVR
- Firm FEI
- 3000302153
- Firm city
- Birmingham
- Firm state
- AL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28