Skip to the content

FDA Medical Device Recalls (openFDA)

Synovis Micro Companies Alliance Inc: GEM 2753 Microvascular Anastomotic Coupler, 2.5 mm, REF GEM2753/I, P/N 5111-00250-010, Rx only, Irradiation Sterilized This is a microsurgical coupler used in the anastomosis of veins and arteries normally encountered in microsurgical vascular reconstructive procedures.

Recall number
Z-2138-2008
Recalling firm
Synovis Micro Companies Alliance Inc
Product
GEM 2753 Microvascular Anastomotic Coupler, 2.5 mm, REF GEM2753/I, P/N 5111-00250-010, Rx only, Irradiation Sterilized This is a microsurgical coupler used in the anastomosis of veins and arteries normally encountered in microsurgical vascular reconstructive procedures.
Status
Terminated
Initiated
2008-05-14
Posted
2008-08-02
Terminated
2010-04-09
Reason
Coupler rings may slip out of the delivery toolcaused by jaws larger than specification.
Root cause
Device Design
Action
Urgent Recall Notification was sent via Fed Ex on May 15, 2008 to 432 hospitals. The letters were addressed to the attention of the Risk Manager. The recall notification instructed the hospital to immediately discontinue use, examine their inventory and, if found, to remove the product and quarantine it for return. A second notification was sent on July 21, 2008 to non-responding consignees.
Quantity
3,812 units
Distribution
Worldwide Distribution --- including USA and countries of Canada, Japan, UAE, Korea, South Africa, Australia, Taiwan, Norway, Netherlands, Sweden, Greece, Norway, Turkey, Portugal, Taiwan, Denmark, Italy, Germany, Japan, Spain, Austria, and Great Britain.
Lot, serial or model codes
Lot numbers: 1808, 1894, 1907, 1923, 1953, 1954, 7139, 7211, 7354, 8998, 8999, 11591, 12364, 15244, 15280, 15281, 18824, 18912, 19294, 19413, 19954, 20263, 20689, 21345, 21515, 21560, 21977, 22556, and 23047
510(k) numbers
["K040163"]
PMA numbers
not on file
Product code
MVR
Firm FEI
3000302153
Firm city
Birmingham
Firm state
AL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register