DJO, LLC: Brand Name: DONJOY Product Name: ICEMAN, W/COLDPAD, NS, RH Model/Catalog Number: 11-9099 Software Version: No Product Description: 11-9099 - ICEMAN, W/COLDPAD, NS, RH The IceMan CLASSIC3 cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No

Recall number
Z-2137-2026
Recalling firm
DJO, LLC
Product
Brand Name: DONJOY Product Name: ICEMAN, W/COLDPAD, NS, RH Model/Catalog Number: 11-9099 Software Version: No Product Description: 11-9099 - ICEMAN, W/COLDPAD, NS, RH The IceMan CLASSIC3 cold therapy unit helps reduce pain and swelling, speeding up rehabilitation. The IceMan helps provide extended cold therapy to aid in a variety of indications and protocols as directed by a medical professional. The IceMan CLASSIC3 utilizes DonJoy's recirculation technology. Component: No
Status
Open, Classified
Initiated
2026-03-19
Posted
2026-05-08
Terminated
not on file
Reason
Due to broken connector which may result in a temporary interruption of prescribed cold therapy.
Root cause
Process control
Action
On 03/19/2026, the firm emailed an "Urgent: Medical Device Recall" Letter to customers to inform them that, twenty-nine complaints have been received reporting a consistent failure mode described as a broken connector. The connector is the component that attaches the end of the circulating hose from the cooler to the pad which is placed on/around the affected limb. Excess water leakage from this component is a key indicator of breakage. Customers are instructed to: Enovis (DJO, LLC) requires customers to take the following actions by 03-31-2026: This notice must be forwarded to all those within their organization or to any organization where the affected product has been transferred. 1. Review stock of the affected products and lot numbers to be returned. Quantities or affected product previously shipped are listed in the table at the opening of this letter. 2. Quarantine these devices from sale. 3. Complete the attached acknowledgment and response form and return to (Customer.Care@enovis.com). 4. To return affected product, please contact (Customer.Care@enovis.com) or call (800) 336-6569 option 4 and customers will be provided with a shipping label to return the affected product. 5. Return all affected products using the shipping label provided. For questions - contact Kenneth Newman at productsafety@enovis.com or call/text 682-299-9133.
Quantity
96
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE. NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The country of Canada.
Lot, serial or model codes
Model/Catalog Number: 11-9099 Lot Code: Lot: 110525 GTIN: 00888912000376
510(k) numbers
not on file
PMA numbers
not on file
Product code
ILO
Firm FEI
3017264244
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register