Integra LifeSciences Corp.: CODMAN EVD BACTISEAL 1.9mm I.D. Catheter Set- indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume. Product Number: 821749

Recall number
Z-2137-2024
Recalling firm
Integra LifeSciences Corp.
Product
CODMAN EVD BACTISEAL 1.9mm I.D. Catheter Set- indicated for gaining access to the ventricles of the brain and can be used with dimensionally compatible devices for draining cerebrospinal fluid (CSF) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume. Product Number: 821749
Status
Open, Classified
Initiated
2024-05-07
Posted
2024-06-12
Terminated
not on file
Reason
Defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the seal may compromise the product sterility.
Root cause
Process control
Action
Integra issued 1st Notification- Urgent: Voluntary Medical Device Recall Letter courier service, facsimile or email between May 7th and May 9th, 2024. Letter states reason for recall, health risk and action to take: Actions to be Taken by Customers: 1. If you do have units of the affected product listed in Table 1, remove them immediately from service and quarantine them. 2. If you do have affected product, check the box on the enclosed Acknowledgement Form I do have affected product. Record the total quantity of the affected product that you have. 3. Forward this notice to those who utilize the device so they are aware of this recall and can identify any affected devices that may remain in clinical areas. 4. If you do not have affected product, check the box, I do not have affected product. 5. Complete the rest of the Acknowledgement Form and return to FCA3@integralife.com or FAX to 1-609-750-4220. 6. Keep a copy of the form for your records. 7. When your form is received, and it is noted that you have affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the affected product. You can request a credit for the quantities returned. Should you have any questions regarding these instructions, please contact Customer Service: Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: custsvcnj@integralife.com
Quantity
106 units
Distribution
Worldwide - US Nationwide distribution in the state of TX and the countries of Canada, China, Croatia, Hungary, Russia, Singapore, Slovenia, South Africa, Taiwan (Province of China), Turkey
Lot, serial or model codes
UDI-DI: 10381780518440 Lot numbers: 7314912, 7320010, 7322933
510(k) numbers
["K233448"]
PMA numbers
not on file
Product code
JXG
Firm FEI
3003418325
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register