Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Vascular, Inc.: Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Recall number
Z-2135-2021
Recalling firm
Medtronic Vascular, Inc.
Product
Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.
Status
Open, Classified
Initiated
2021-06-07
Posted
not on file
Terminated
not on file
Reason
During stent graft deployment, the radiopaque (RO) marker bond detached from the distal end of the graft cover.
Root cause
Under Investigation by firm
Action
On 06/07/2021, the firm initiated a verbal notification regarding an "URGENT MEDICAL DEVICE RECALL" which informed customers that Earlier this week, Medtronic identified the potential for a specific subset of 20Fr Endurant II/IIs Stent Graft Systems to have the potential for the Radiopaque Marker to detach during deployment. There is evidence that devices that used one batch of graft cover subassemblies has a higher potential to exhibit this failure mode. This is an acute failure of the delivery system, there are no concerns with previously implanted stent grafts. We are requesting that you immediately refrain from using this specific subset of products and place it in quarantine until further notice. You will be receiving an official recall notice with instructions for product return for credit or replacement. Your Sales Representative will assist you with this process. We apologize for the inconvenience; however, this action is necessary to ensure the utmost patient safety with the use of our products. On 06/18/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification Letter via UPS 2-Day delivery that restated the verbal notification. The Customer Notification Letter instructed customers to: 1)Return all unused affected devices via their sales representative or calling customer service at 1-888-283-7868 or email rs.aorticorders@medtronic.com. 2)Complete the enclosed Customer Confirmation Form and email to rs.cfqfca@medtronic.com. 3)Forward this notice to all those who need to be aware within their organization. For questions or assistance, contact Customer Service at 1-888-283-7868.
Quantity
18 devices
Distribution
U.S.: GA, MA, MD, NY, SC, TX, WA, and WV O.U.S.: Argentina, Australia, Germany, Greece, Israel, Italy, Netherlands, Poland, South Africa, Spain, Switzerland and United Kingdom.
Lot, serial or model codes
Endurant II Stent Graft System: Model (REF): ETCF3636C49EE GTIN: 00763000099626 Serial Numbers: V29778214; V29778215; Model (REF): ETUF3214C102EE GTIN: 00643169780958 Serial Numbers: V29775791; V29775792; V29775793; V29775794; V29775795; Model (REF): ETBF3216C166EE GTIN: 00763000099626 Serial Numbers: V29778214; V29778215; Model (REF): ETUF3614C102E GTIN: 00643169232631 Serial Numbers: V29781886 V29781883 V29781889 V29781887 V29781890 V29781891 V29781882
510(k) numbers
not on file
PMA numbers
["P100021"]
Product code
MIH
Firm FEI
3005364322
Firm city
Santa Rosa
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register