FDA Medical Device Recalls (openFDA)
MOTEK MEDICAL B.V.: Body Weight Support (BWS) Light; Model No. RB011; a component of C-Mill, Grail, M-Gait - Product Usage: is designed to facilitate functional gait training of subjects with gait and/or balance impairment by providing dynamic partial vertical unloading during walking on a treadmill. The device will be used as an assistive device for gait deficiency rehabilitation and research purposes.
- Recall number
- Z-2135-2020
- Recalling firm
- MOTEK MEDICAL B.V.
- Product
- Body Weight Support (BWS) Light; Model No. RB011; a component of C-Mill, Grail, M-Gait - Product Usage: is designed to facilitate functional gait training of subjects with gait and/or balance impairment by providing dynamic partial vertical unloading during walking on a treadmill. The device will be used as an assistive device for gait deficiency rehabilitation and research purposes.
- Status
- Open, Classified
- Initiated
- 2019-07-29
- Posted
- not on file
- Terminated
- not on file
- Reason
- The firm discovered a potential manufacturing deviation of the yoke of the Body Weight Support (BWS) Light that in rare occasions may result in the yoke detaching from the cable guide.
- Root cause
- Process control
- Action
- On 07/29/2019 the firm notified their customers of the Field Safety Notice: BWS-L Yoke 2019 letter sent via email. The firm informed their customers to share the notice with all those within their organizations who need to be aware. Customers were asked to confirm receipt of the Field Safety Notice. Additionally, customers were asked to reply by email to support@motekforcelink.com and attach a completed and signed copy of the Advisory Notice Confirmation (last page of the notice) within 10 business days of receipt of the notice. The firm provided the following information corrective and/or prevention action: "In all cases, if you are treating patients with cranial deficiencies, stop using the system for these patients immediately". Three possible options were provided. Customers were asked to select one that is most appropriate for their institute and send them this information via the Advisory Notice Confirmation: A. You request a Yoke Replacement Kit to be sent to you directly, such that a skilled person from your institute can do the replacement under the firm's remote supervision. The replacement consists of removing two easily accessible bolts, replacing the affected components, and reinserting two new bolts. Verification will require to take two pictures of the reinserted bolts, and confirmation by the firm is needed before you can use the system again for patients with cranial deficiencies. Detailed instructions will be provided. B. Or, you do not treat patients with cranial deficiencies, and do not want to order a Yoke Replacement Kit. If so, you can continue using the system as normal and can consider this Field Safety Notice to be not applicable for your usage (even though we still need you to send in the Advisory Notice Confirmation). In the next service visit, we will replace the yoke to further minimize any risk. C. Or, you do treat patients with cranial deficiencies, but you do not want to order a Yoke Replacement Kit. In this case, you
- Quantity
- 80 units
- Distribution
- Worldwide distribution - US Nationwide distributions including in the states of AZ, CA, FL, KY, MA, MI, NJ, NY, NC, and OH. The countries of Argentina, Austria, Belgium, China, Curacao, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Japan, Mexico, Netherlands, Russia, South Korea, Spain, Switzerland, and United Arab Emirates.
- Lot, serial or model codes
- AS9924-00-0002, RB011-00-0005,RB011-00-0006, RB011-00-0010, RB011-00-0012, RB011-00-0015, RB011-00-0017, RB011-00-0018, RB011-00-0019, RB011-00-0020, RB011-00-0021, RB011-0022, RB011-0023, RB011-0024, RB011-0025, RB011-0026, RB011-0027, RB011-0028, RB011-0029, RB011-0030, RB011-0031, RB011-0032, RB011-0033, RB011-0034, RB011-0035, RB011-0036, RB011-0037, RB011-0038, RB011-0039, RB011-0040, RB011-0041, RB011-0042, RB011-0044, RB011-0045, RB011-0046, RB011-0047, RB011-0048, RB011-0049, RB011-0050, RB011-0051, RB011-0052, RB011-0053, RB011-0054, RB011-0055, RB011-0056, RB011-0057, RB011-0058, RB011-0059, RB011-0060, RB011-0061, RB011-0062, RB011-0063, RB011-0064, RB011-0065, RB011-0066, RB011-0067, RB011-0068, RB011-0069, RB011-0070, RB011-0071, RB011-0072, RB011-0073, RB011-0074, RB011-0075, RB011-0076, RB011-0077, RB011-0078, RB011-0079, RB011-0080, RB011-0081, RB011-0082, RB011-0083, RB011-0084, RB011-0085, RB011-0086, RB011-0087, RB011-0088, RB011-0089, RB011-0090, RB011-0091, and RB011-0092.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- BXB
- Firm FEI
- 3006750942
- Firm city
- Amsterdam
- Firm state
- not on file
- Firm country
- Netherlands
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28