Skip to the content

FDA Medical Device Recalls (openFDA)

Acumed LLC: Smooth or threaded metallic bone fixation fastener. 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in part, "Congruent Plate System... Size: 3.5mm x 14.0 mm Locking Cortical Screw... REF COL-3140-S LOT W46772... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR ". Product is used as an orthopedic implant.

Recall number
Z-2135-2008
Recalling firm
Acumed LLC
Product
Smooth or threaded metallic bone fixation fastener. 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in part, "Congruent Plate System... Size: 3.5mm x 14.0 mm Locking Cortical Screw... REF COL-3140-S LOT W46772... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR ". Product is used as an orthopedic implant.
Status
Terminated
Initiated
2007-07-09
Posted
2008-09-11
Terminated
2010-02-03
Reason
Product mislabeled. The inner sticky labels for the patient identified the product incorrectly as a Cortical screw (non-locking). The product package was labeled correctly.
Root cause
Packaging process control
Action
The firm issued a Product Removal notification on 8-9-2007. The notification was reportedly sent by either FAX or e-mail. The firm notified consignees of the issue and gave the consignees the choice of returning the screws. The notification did not include any instructions for a sub-recall. If you have any questions contact Jason Gozdecki at 1-888-627-9957.
Quantity
400 screws
Distribution
Class III Recall - Worldwide Distribution --- USA including countries of UK and Japan.
Lot, serial or model codes
Part number: COL-3140-S; Work order number/Lot: W46772 (W=work order, ###=sequential number); Expiration date: 2012-02 (Year/Month)
510(k) numbers
["K012655","K942340","K942341"]
PMA numbers
not on file
Product code
HWC
Firm FEI
1000125930
Firm city
Hillsboro
Firm state
OR
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register