FDA Medical Device Recalls (openFDA)
Acumed LLC: Smooth or threaded metallic bone fixation fastener. 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in part, "Congruent Plate System... Size: 3.5mm x 14.0 mm Locking Cortical Screw... REF COL-3140-S LOT W46772... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR ". Product is used as an orthopedic implant.
- Recall number
- Z-2135-2008
- Recalling firm
- Acumed LLC
- Product
- Smooth or threaded metallic bone fixation fastener. 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in 3.5mm x 14.0 mm Locking Cortical Screw. Product labeled in part, "Congruent Plate System... Size: 3.5mm x 14.0 mm Locking Cortical Screw... REF COL-3140-S LOT W46772... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR ". Product is used as an orthopedic implant.
- Status
- Terminated
- Initiated
- 2007-07-09
- Posted
- 2008-09-11
- Terminated
- 2010-02-03
- Reason
- Product mislabeled. The inner sticky labels for the patient identified the product incorrectly as a Cortical screw (non-locking). The product package was labeled correctly.
- Root cause
- Packaging process control
- Action
- The firm issued a Product Removal notification on 8-9-2007. The notification was reportedly sent by either FAX or e-mail. The firm notified consignees of the issue and gave the consignees the choice of returning the screws. The notification did not include any instructions for a sub-recall. If you have any questions contact Jason Gozdecki at 1-888-627-9957.
- Quantity
- 400 screws
- Distribution
- Class III Recall - Worldwide Distribution --- USA including countries of UK and Japan.
- Lot, serial or model codes
- Part number: COL-3140-S; Work order number/Lot: W46772 (W=work order, ###=sequential number); Expiration date: 2012-02 (Year/Month)
- 510(k) numbers
- ["K012655","K942340","K942341"]
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 1000125930
- Firm city
- Hillsboro
- Firm state
- OR
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as