Skip to the content

FDA Medical Device Recalls (openFDA)

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Azurion 7 M20. Fluoroscopic X-Ray System.

Recall number
Z-2134-2025
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Azurion 7 M20. Fluoroscopic X-Ray System.
Status
Open, Classified
Initiated
2025-06-04
Posted
2025-07-18
Terminated
not on file
Reason
Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
Root cause
Device Design
Action
A "Device URGENT Medical Device Correction" notification dated 6/03/2025 was mailed to consignees. Consignees are instructed to circulate the provided notification to all system users and keep a copy of the recall notification with affected devices to maintain awareness. If you notice unusual noise during C-arm roll motion, or if the C-arm roll motion is disabled and the user message Motorized movement is not available appears, immediately contact a Philips representative. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. Affected systems may continue to be used in accordance with their Instructions for Use (IFU). Complete and return the response form included in this Urgent Medical Device letter. Consignees are provided instructions to avoid rapidly pressing and releasing footswitch pedals and how to restart systems that experience the failure mode. The firm will deploy a software update in all affected system by June 2025. Consignees with any questions should contact their local Philips representative at 1-800-722-9377.
Quantity
78 units
Distribution
Worldwide distribution: US (nationwide) to states of: CA, CO, DC, DE, FL, IL, IN, KY, MA, MD, MN, MO, MT, NJ, NV, NY, OH, SC, TX, WA; and OUS (Foreign) to countries of: Argentina (AR), Australia (AU), Canada (CA), Chile (CL), Germany (DE), Dominican Republic (DO), Egypt (EG), ES, France (FR), United Kingdom (GB), Italy (IT), Japan (JP), Korea, Republic (KR), Mexico (MX), Netherlands (NL), New Zealand (NZ), Oman (OM), Reunion (RE), Saudi Arabi (SA), Thailand (TH), Viet Nam (VN).
Lot, serial or model codes
System Model No. 722079; UDI: 00884838085268; Serial No. 525, 584, 587, 613, 631, 658, 708, 774, 777, 793, 800, 801, 838, 842, 843, 862, 865, 866, 879, 881, 896, 899, 900, 905, 908, 909, 912, 915, 923, 925, 930, 939, 948, 955, 957, 992, 750, 814, 321, 370, 549, 567, 670, 687, 688, 693, 707, 743, 759, 767, 768, 775, 785, 786, 791, 819, 855, 871, 872, 902, 916, 917, 933, 936, 956, 960, 963, 965, 966, 969, 977, 993.
510(k) numbers
["K163715","K172822","K181830"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register