Skip to the content

FDA Medical Device Recalls (openFDA)

Karl Storz Endoscopy: Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Video Mediastinoscope

Recall number
Z-2134-2016
Recalling firm
Karl Storz Endoscopy
Product
Image1 SPIES X-LINK Camera Control Unit (CCU), Product Model No. TC301, when used with Image1 SPIES D1 camera head, Product Model No. TH116 & TH115, and DC1 Video Mediastinoscope
Status
Terminated
Initiated
2016-06-20
Posted
2016-07-06
Terminated
2017-01-05
Reason
Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).
Root cause
Device Design
Action
KARL STORZ sent an Customer Notification letter dated June 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. KARL STORZ has become aware become aware of a potential electrical interference issue when the lmage1 SPIES X-LINK and lmage1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinoscope and an electro-surgery generator (ESG). Customers are instructed to complete the attached acknowledgement Form and to return it to KARL STORZ. Customers with questions are instructed to contact Technical Support Dept. at (800) 421-0837, press 2, and then press 1.
Quantity
619 units total (90 units in US)
Distribution
Worldwide Distribution - US Distribution to the states of : IL, NM, TX, MD, TN, PA, FL, KY, MO, NY, MN, DC, MA, WI, CT, MI, ID, CO, GA, NJ and AR., and to the countries of : Argentina, Australia, Belgium, Bulgaria, China, Germany, Finland, Great Britain, India, Italy and Japan.
Lot, serial or model codes
NT743473-P NT743472-P NT742840-P XU738911-P YU741045-P XU738248-P WU737260-P PV715203-P PV715207-P NT742829-P XU738899-P WU737275-P YV724235-P WU737269-P WU737268-P VU736411-P XU738901-P XU738910-P ZU742677-P YU741050-P VU736405-P TU733422-P TU733419-P SU733063-P VU736390-P VU736386-P UU734392-P RV716122-P OT744400-P NT743287-P OT744395-P NU727551-P NU727552-P NU727706-P TU734218-P TU734217-P NT742846-P YU741046-P OU728667-P TU733735-P VU736388-P NT742848-P NT742838-P OU728672-P OU728427-P WU737813-P OU729135-P OU728431-P OU728430-P WU737805-P ZU742670-P ZU742671-P YU741049-P YU740475-P ZU742681-P RV716116-P SU733055-P SU733057-P SU733059-P SU733060-P TU733418-P TU733420-P TU733421-P TU733423-P TU733424-P TU733425-P TU733427-P WU737810-P XU738914-P YU740004-P YU740471-P YU740479-P YU740480-P YU741036-P NV711744-P PV715216-P OU728670-P NU726430-P OT744399-P OT744389-P OT744386-P YU739999-P YU740003-P YU740005-P UU734395-P YV724443-P YV724442-P YV724440-P YV724438-P YV724245-P
510(k) numbers
["K131953"]
PMA numbers
not on file
Product code
FET
Firm FEI
3007475226
Firm city
El Segundo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register