Skip to the content

FDA Medical Device Recalls (openFDA)

DePuy Spine, Inc.: CONFIDENCE SPINAL CEMENT SYSTEM KIT, Product Code: 2839-13-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures.

Recall number
Z-2134-2013
Recalling firm
DePuy Spine, Inc.
Product
CONFIDENCE SPINAL CEMENT SYSTEM KIT, Product Code: 2839-13-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures.
Status
Terminated
Initiated
2013-04-17
Posted
2013-09-03
Terminated
2014-08-13
Reason
During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.
Root cause
Process control
Action
DePut Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated April 15, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the con signee. Consignees were instructed to review their inventory and quarantine for return any products listed in the recall notice., complete the enclosed Business Reply Form, indicating level of inventory of the affected product and return the form to Depuy via fax to 888-943-4897or email to DePuy7862@stericycle.com and to return the product with the enclosed UPS label to Stericycle. For questions regarding this recall call 508-977-6606.
Quantity
3896 units (Overall quantity distributed appears greater than quantities manufactured due to the distribution pathways)
Distribution
Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK
Lot, serial or model codes
HPBB0B, HPBBWM, HPCB75, HPCCBN, HPBB0R, HPBBWP, HPCB7D, HPDBBV, HPBB28, HPBBZ7, HPCB7F, HPDBCG, HPBB2D, HPBBZF, HPCB7T, HPDBD1, HPBB2F, HPBCKH, HPCB7V, HPDBD5, HPBB4R, HPBDL9, HPCB8G, HPDBF9, HPBBV1, HPCB67, HPCBLC, HPDBGH, HPBBWG, HPCB68, HPCBM1, HPDBGK, HPBBWK, HPCB74, HPCCBM, HPDBGL
510(k) numbers
["K060300"]
PMA numbers
not on file
Product code
NDN
Firm FEI
1526439
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register