FDA Medical Device Recalls (openFDA)
DePuy Spine, Inc.: CONFIDENCE SPINAL CEMENT SYSTEM KIT, Product Code: 2839-13-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures.
- Recall number
- Z-2134-2013
- Recalling firm
- DePuy Spine, Inc.
- Product
- CONFIDENCE SPINAL CEMENT SYSTEM KIT, Product Code: 2839-13-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures.
- Status
- Terminated
- Initiated
- 2013-04-17
- Posted
- 2013-09-03
- Terminated
- 2014-08-13
- Reason
- During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.
- Root cause
- Process control
- Action
- DePut Synthes Spine sent an Urgent Voluntary Product Recall Notification letter dated April 15, 2013, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the con signee. Consignees were instructed to review their inventory and quarantine for return any products listed in the recall notice., complete the enclosed Business Reply Form, indicating level of inventory of the affected product and return the form to Depuy via fax to 888-943-4897or email to DePuy7862@stericycle.com and to return the product with the enclosed UPS label to Stericycle. For questions regarding this recall call 508-977-6606.
- Quantity
- 3896 units (Overall quantity distributed appears greater than quantities manufactured due to the distribution pathways)
- Distribution
- Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK
- Lot, serial or model codes
- HPBB0B, HPBBWM, HPCB75, HPCCBN, HPBB0R, HPBBWP, HPCB7D, HPDBBV, HPBB28, HPBBZ7, HPCB7F, HPDBCG, HPBB2D, HPBBZF, HPCB7T, HPDBD1, HPBB2F, HPBCKH, HPCB7V, HPDBD5, HPBB4R, HPBDL9, HPCB8G, HPDBF9, HPBBV1, HPCB67, HPCBLC, HPDBGH, HPBBWG, HPCB68, HPCBM1, HPDBGK, HPBBWK, HPCB74, HPCCBM, HPDBGL
- 510(k) numbers
- ["K060300"]
- PMA numbers
- not on file
- Product code
- NDN
- Firm FEI
- 1526439
- Firm city
- Raynham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28