Xhale Assurance, Inc.: Nasal Alar SpO2 Sensor, Adult/Pediatric SpO2 Sensor, Rx Only Product Usage: The Assurance Nasal Alar SpO2 Sensor is indicated for single patient use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients, (at least 4 years and older and weighing > 15kg). The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision

Recall number
Z-2133-2019
Recalling firm
Xhale Assurance, Inc.
Product
Nasal Alar SpO2 Sensor, Adult/Pediatric SpO2 Sensor, Rx Only Product Usage: The Assurance Nasal Alar SpO2 Sensor is indicated for single patient use for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients, (at least 4 years and older and weighing > 15kg). The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision
Status
Terminated
Initiated
2019-05-08
Posted
not on file
Terminated
2022-04-25
Reason
These Instructions for Use has missing instructions related to the checking and changing of the application site procedure.
Root cause
Device Design
Action
Xhale Assurance, Inc. notified consignees on about 05/08/2019, via "URGENT - Medical Device Correction" letter. Customers were instructed to insert the IFU Addendum into each case of sensors for ready reference, review all information provided in the updated IFU Addendum with all members of your staff whom need to be aware of the updated instructions, and complete and return the provided Customer Reply Form. Distributors were instructed to insert the IFU Addendum into each case of sensors for ready reference, provide a copy of the Field Safety Notice and IFU Addendum to any customer to whom affected devices have been distributed to, and complete and return the provided Customer Reply Form.
Quantity
701,272 devices total
Distribution
US Nationwide distribution to AZ, CA, FL, KS, LA, MA , MO, MS, NC, NY, OH, PA, TX, WI. International distribution to Canada, Germany, Malaysia, United Kingdom.
Lot, serial or model codes
Reference Codes (Lot Numbers): 0303 (101619, 101624, 101628, 101632, 101637, 101642, 101646, 101647, 101651, 101701, 201651, 101707, 101715, XH101711, XH101720, XH101715, XH101726, XH101729, XH101733, XH101737, XH101820, XH101842), 301-11214 (TR101711, TR101715, TR101720, TR101726, TR101729, TR101744, TR101733, TR101737, TR101748, TR101742, TR101746, TR101750, TR101803, TR101811, TR101817, TR101820, TR101824, TR101829, TR101833, TR101838, TR101842, TR101847, TR101852), 0201-A01(PF101736, PF101806, PF101814, PF101817, PF101819, PF101822, PF101828, PF101833, PF101838). Instructions for Use for revisions 10412_7 and prior (i.e. 10412_6, 10412_5, etc) as well as 10358_6 and prior.  Shelf life is 3 years for sensors manufactured prior to Oct 2018. 5 years for sensor manufactured on or after Oct 2018.
510(k) numbers
["K143216","K171423"]
PMA numbers
not on file
Product code
DQA
Firm FEI
1000113495
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register