Skip to the content

FDA Medical Device Recalls (openFDA)

GE Medical Systems, LLC: Discovery MR750w 3.0T

Recall number
Z-2132-2025
Recalling firm
GE Medical Systems, LLC
Product
Discovery MR750w 3.0T
Status
Open, Classified
Initiated
2025-05-23
Posted
2025-07-18
Terminated
not on file
Reason
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Root cause
Device Design
Action
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 05/23/2025 delivered using traceable means. The notice explained the issue, risk, and requested the following: Actions to be taken by Customer/User: You can continue to use your MR system by following the instructions below: Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to recall.60999@gehealthcare.com.
Quantity
not on file
Distribution
Worldwide distribution - US Nationwide and the country of China.
Lot, serial or model codes
GTIN: 00840682103817 00195278229519
510(k) numbers
not on file
PMA numbers
not on file
Product code
LNH
Firm FEI
2183553
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register