FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUMBER: REF 74035822 SIZE 3, CATALOG NUMBER: REF 74035823 SIZE 4, CATALOG NUMBER: REF 74035824 SIZE 5, CATALOG NUMBER: REF 74035825 SIZE 6, CATALOG NUMBER: REF 74035826 SIZE 7, CATALOG NUMBER: REF 74035827 SIZE 8, CATALOG NUMBER: REF 74035828 SIZE 9, CATALOG NUMBER: REF 74035829 SIZE 10, CATALOG NUMBER: REF 74035830

Recall number
Z-2132-2020
Recalling firm
Smith & Nephew, Inc.
Product
JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of motion trialing to assess joint tension and component alignment during surgery. The provisional moveable spikes (spike rail) aids in provisional fixation of the trial. SIZE 1, CATALOG NUMBER: REF 74035821 SIZE 2, CATALOG NUMBER: REF 74035822 SIZE 3, CATALOG NUMBER: REF 74035823 SIZE 4, CATALOG NUMBER: REF 74035824 SIZE 5, CATALOG NUMBER: REF 74035825 SIZE 6, CATALOG NUMBER: REF 74035826 SIZE 7, CATALOG NUMBER: REF 74035827 SIZE 8, CATALOG NUMBER: REF 74035828 SIZE 9, CATALOG NUMBER: REF 74035829 SIZE 10, CATALOG NUMBER: REF 74035830
Status
Terminated
Initiated
2020-01-28
Posted
not on file
Terminated
2023-12-11
Reason
The spike rail was manufactured with an oversized thru-hole which could potentially cause the dowel pin to fall out and the device to disassemble.
Root cause
Process design
Action
The firm initiated the recall by email on 01/28/2020. The initial consignee notified the district sales representatives on 01/28/2020 by email. The district sales representatives checked their inventory and the inventory of the hospital they service.
Quantity
200 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
SIZE 1, CATALOG NUMBER: REF 74035821; LOT/BATCH CODES: 19EM19112R, 19FM05587R, 19FM11174R, 19HM18075, 19HM19648, 19HM21015, 19HM21016, 19HM23242  SIZE 2, CATALOG NUMBER: REF 74035822; LOT/BATCH CODES: 19EM21573R, 19EM21573S, 19FM05589R, 19HM21031, 19HM21033, 19HM21878, 19HM21881  SIZE 3, CATALOG NUMBER: REF 74035823; LOT/BATCH CODES: 19EM21578R, 19HM19009, 19JM09476, 19JM18509, 19JM18525, 19JM21072, 19KM14473, 19KM17780  SIZE 4, CATALOG NUMBER: REF 74035824; LOT/BATCH CODES: 19EM14365R, 19EM14365S, 19FM11173, 19HM18078, 19HM19659, 19HM21861, 19JM21075, 19KM15573, 19KM1557  SIZE 5, CATALOG NUMBER: REF 74035825; LOT/BATCH CODES: 19EM14367R, 19HM18079, 19HM21855, 19KM12683   SIZE 6, CATALOG NUMBER: REF 74035826; LOT/BATCH CODES: 19EM24480R, 19HM18080, 19HM21856, 19KM21512  SIZE 7, CATALOG NUMBER: REF 74035827; LOT/BATCH CODES: 19EM14371R, 19FM01063R, 19FM11177R, 19HM18335, 19HM21873, 19HM21874, 19KM17781  SIZE 8, CATALOG NUMBER: REF 74035828; LOT/BATCH CODES: 19EM24498R, 19EM24498S, 19FM05591R, 19FM11178R, 19FM11178S, 19JM07464, 19JM07465, 19KM07278, 19KM12699  SIZE 9, CATALOG NUMBER: REF 74035829; LOT/BATCH CODES: 19EM21582R, 19FM11179R, 19GM21787, 19JM07467, 19KM07276, 19KM15575, 19KM15576  SIZE 10, CATALOG NUMBER: REF 74035830; LOT/BATCH CODES: 19EM14372R, 19FM11180, 19HM18254, 19JM07468, 19JM07470, 19JM07472, 19KM05610, 19KM15571
510(k) numbers
["K190085"]
PMA numbers
not on file
Product code
HSX
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register