ZOLL Medical Corporation: ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference Guide (QRGs) as follows: Part Numbers: 1. 799-EGL2-02 2. 799-EGL2-02-01 3. 799-EGL2-02-04 4. 799-EGL2-02-05 5. 799-EGL2-02-08 6. 799-EGL2-02-10 7. 799-EGL2-02-11 8. 799-EGL2-02-15 9. 799-EGL2-02-16 10. 799-EMBP-05-01 11. 799-EMBP-05-02 12. 799-EMBP-05-08 13. 799-EMBP-05-10 14. 799-EMBP-05-11 15. 799-EMBP-05-12 16. 799-EMBP-05-12-68 17. 799-EMBP-05-16 18. 799-EMBP-05-23 19. 799-EMVP-05 20. 799-EMVP-05-01 21. 799-EMVP-05-05 22. 799-EMVP-05-10 23. 799-EMVP-05-15 24. 799-EMVP-05-16 Manual Part Numbers 906-0731-01; 9650-002365-01 (US); 906-0731-01-05 (CAN); 906-0731-01-10; 9650-0002363-01; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 906-0731-01; 906-0731-01-05 (CAN); 906-0731-01-04; 906-0731-01-05; 906-0731-01-08; 906-0731-01-10; 906-0731-01-11; 906-0731-01-15; 906-0731-01-16; 9650-0002363-01; 9650-0002363-02; 9650-0002363-08; 9650-0002363-10; 9650-0002363-11; 9650-0002363-12; 9650-0002363-16; 9650-0002363-23; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 9650-0002363-10; 9650-0002363-15; 9650-0002363-16 Quick Reference Guide (QRG) Part Numbers: 907-0731-04; 9652-000499-01; 9652-000511-01; 907-0731-04; 9652-000511-01; 907-0731-04; 9652-000499-01; 9652-000499-04; 9652-000499-08; 9652-000499-10; 9652-000499-11; 907-0731-04; 9652-000499-16; 9652-000511-01; 9652-000511-02; 9652-000511-08; 9652-000511-10; 9652-000511-11; 9652-000511-12; 9652-000511-16; 9652-000511-23; 907-0731-04; 9652-000511-01; 9652-000511-05; 9652-000511-10; 9652-000511-15; 9652-000511-16

Recall number
Z-2130-2024
Recalling firm
ZOLL Medical Corporation
Product
ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference Guide (QRGs) as follows: Part Numbers: 1. 799-EGL2-02 2. 799-EGL2-02-01 3. 799-EGL2-02-04 4. 799-EGL2-02-05 5. 799-EGL2-02-08 6. 799-EGL2-02-10 7. 799-EGL2-02-11 8. 799-EGL2-02-15 9. 799-EGL2-02-16 10. 799-EMBP-05-01 11. 799-EMBP-05-02 12. 799-EMBP-05-08 13. 799-EMBP-05-10 14. 799-EMBP-05-11 15. 799-EMBP-05-12 16. 799-EMBP-05-12-68 17. 799-EMBP-05-16 18. 799-EMBP-05-23 19. 799-EMVP-05 20. 799-EMVP-05-01 21. 799-EMVP-05-05 22. 799-EMVP-05-10 23. 799-EMVP-05-15 24. 799-EMVP-05-16 Manual Part Numbers 906-0731-01; 9650-002365-01 (US); 906-0731-01-05 (CAN); 906-0731-01-10; 9650-0002363-01; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 906-0731-01; 906-0731-01-05 (CAN); 906-0731-01-04; 906-0731-01-05; 906-0731-01-08; 906-0731-01-10; 906-0731-01-11; 906-0731-01-15; 906-0731-01-16; 9650-0002363-01; 9650-0002363-02; 9650-0002363-08; 9650-0002363-10; 9650-0002363-11; 9650-0002363-12; 9650-0002363-16; 9650-0002363-23; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 9650-0002363-10; 9650-0002363-15; 9650-0002363-16 Quick Reference Guide (QRG) Part Numbers: 907-0731-04; 9652-000499-01; 9652-000511-01; 907-0731-04; 9652-000511-01; 907-0731-04; 9652-000499-01; 9652-000499-04; 9652-000499-08; 9652-000499-10; 9652-000499-11; 907-0731-04; 9652-000499-16; 9652-000511-01; 9652-000511-02; 9652-000511-08; 9652-000511-10; 9652-000511-11; 9652-000511-12; 9652-000511-16; 9652-000511-23; 907-0731-04; 9652-000511-01; 9652-000511-05; 9652-000511-10; 9652-000511-15; 9652-000511-16
Status
Open, Classified
Initiated
2024-04-30
Posted
2024-06-20
Terminated
not on file
Reason
Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy
Root cause
Under Investigation by firm
Action
Zoll issued Urgent Corrective Action Letter dated 4/22/24. Letter states reason for recall, health, risk and action to take: Customers who have affected manuals should take the following steps: (1) Locate all old versions of affected Operator s Guides and/or Quick Reference Guides. (2) Complete and return the attached customer response form via e-mail, fax, or mail to request replacement manuals. (3) Alert all users of the ZOLL Ventilator - with MRI Compatibility, to ensure they are aware of the safe distance requirements and are trained to these manual updates. (4) Dispose of obsolete guides. Technical support number are 1-800-348-9011 or 1-978-421-9460 and are available to assist users with any aspect of this notice
Quantity
2361 units
Distribution
Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Argentina, Australia, Azerbaijan, Bangladesh, Belgium, Bahrain, Bolivia, Brazil, Canada, Switzerland, China, Colombia, Czech Republic, Germany, Spain, Great Britain (UK), Greece, Guatemala, Guam, Hong Kong, Croatia, Hungary, Indonesia, Israel, India, Iraq, Italy, Japan, South Korea, Lebanon, Myanmar, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Thailand, Turkey, Taiwan, Vietnam.
Lot, serial or model codes
Part Numbers/UDI Zoll 731 Defibrillators: 1. 799-EGL2-02 00847946020910 2. 799-EGL2-02-01 00847946023454 3. 799-EGL2-02-04 00847946036867 4. 799-EGL2-02-05 00847946036874 5. 799-EGL2-02-08 6. 799-EGL2-02-10 00847946036881 7. 799-EGL2-02-11 8. 799-EGL2-02-15 00847946050665 9. 799-EGL2-02-16 10. 799-EMBP-05-01 00847946024185 11. 799-EMBP-05-02 12. 799-EMBP-05-08 13. 799-EMBP-05-10 14. 799-EMBP-05-11 15. 799-EMBP-05-12 00847946036959 16. 799-EMBP-05-12-68 17. 799-EMBP-05-16 18. 799-EMBP-05-23 19. 799-EMVP-05 00847946020767 20. 799-EMVP-05-01 00847946023430 21. 799-EMVP-05-05 00847946034351 22. 799-EMVP-05-10 23 799-EMVP-05-15 00847946049850 24 799-EMVP-05-16 Manual Part Numbers 906-0731-01; 9650-002365-01 (US); 906-0731-01-05 (CAN); 906-0731-01-10; 9650-0002363-01; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 906-0731-01; 906-0731-01-05 (CAN); 906-0731-01-04; 906-0731-01-05; 906-0731-01-08; 906-0731-01-10; 906-0731-01-11; 906-0731-01-15; 906-0731-01-16; 9650-0002363-01; 9650-0002363-02; 9650-0002363-08; 9650-0002363-10; 9650-0002363-11; 9650-0002363-12; 9650-0002363-16; 9650-0002363-23; 906-0731-01 (Commercial); 906-0731-03 (Military); 9650-0002363-01; 9650-0002363-05; 9650-0002363-10; 9650-0002363-15; 9650-0002363-16 Quick Reference Guide (QRG) Part Numbers: 907-0731-04; 9652-000499-01; 9652-000511-01; 907-0731-04; 9652-000511-01; 907-0731-04; 9652-000499-01; 9652-000499-04; 9652-000499-08; 9652-000499-10; 9652-000499-11; 907-0731-04; 9652-000499-16; 9652-000511-01; 9652-000511-02; 9652-000511-08; 9652-000511-10; 9652-000511-11; 9652-000511-12; 9652-000511-16; 9652-000511-23; 907-0731-04; 9652-000511-01; 9652-000511-05; 9652-000511-10; 9652-000511-15; 9652-000511-16
510(k) numbers
["K162832","K111473"]
PMA numbers
not on file
Product code
CBK
Firm FEI
1220908
Firm city
Chelmsford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register