Skip to the content

FDA Medical Device Recalls (openFDA)

Custom Ultrasonics, Inc.: The System 83 Plus, Endoscope Washer/Disinfector Designed for the simultaneous reprocessing of up to two flexible submersible endoscopes that are used in the gastrointestinal and/or pulmonary tracts.

Recall number
Z-2130-2016
Recalling firm
Custom Ultrasonics, Inc.
Product
The System 83 Plus, Endoscope Washer/Disinfector Designed for the simultaneous reprocessing of up to two flexible submersible endoscopes that are used in the gastrointestinal and/or pulmonary tracts.
Status
Terminated
Initiated
2016-05-06
Posted
2016-07-06
Terminated
2017-04-18
Reason
To warn customers that Custom Ultrasonics, Inc. System 83 Plus AERs should not be used for cleaning and/or high-level disinfection of duodenoscopes until further notice.
Root cause
Process control
Action
Custom Ultrasonics, Inc. sent an Urgent Medical Device Recall letter dated May 6, 2016, to all affected customers. The letter advised customers not to use the System 83 Plus AERs for cleaning and/or high-level disinfection of duodenoscopes until further notice. In addition, they sent a warning label for customers to affix in a prominent location on the System 83 Plus AER's on or before June 3, 2016 stating the following: "This device is not indicated for reprocessing of duodenoscopes. Do not reprocess any duodenoscopes in this device until further notice. For alternative reprocessing options, please contact the duodenoscope manufacturer." For questions regarding this recall call 215-364-1477.
Quantity
2,345 units
Distribution
Nationwide Distribution to IN, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD., ME, MI, MN, MO,MS, MT, NC, NE, NH, NJ, NM, NV, NY OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT,WA, WI, WV, and WY.
Lot, serial or model codes
All serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
NVE
Firm FEI
2523209
Firm city
Ivyland
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register