FDA Medical Device Recalls (openFDA)
Orthofix, Inc: H-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 12MM Spacer (003-H-12001); b) 15MM Spacer (003-H-15001); 20MM Spacer (003-H-20001); 25MM Spacer (003-H-25001); 30 MM (003-H-30001)
- Recall number
- Z-2128-2018
- Recalling firm
- Orthofix, Inc
- Product
- H-Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) 12MM Spacer (003-H-12001); b) 15MM Spacer (003-H-15001); 20MM Spacer (003-H-20001); 25MM Spacer (003-H-25001); 30 MM (003-H-30001)
- Status
- Terminated
- Initiated
- 2018-04-06
- Posted
- 2018-06-07
- Terminated
- 2021-06-30
- Reason
- Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
- Root cause
- Device Design
- Action
- On 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
- Quantity
- 75
- Distribution
- U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
- Lot, serial or model codes
- Lot: B1212400, Date of Manufacture: 2017-11-13, Use By: 2022-10-13, (UDI)/Lot: a) (18056099647908) B1096740; b) (18056099647915) B1096733; c) (18056099647939) B1096734; d) (18056099647953) B1096735; e) (18056099647960) B1100226
- 510(k) numbers
- ["K142135"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 2183449
- Firm city
- Lewisville
- Firm state
- TX
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28