Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.

Recall number
Z-2127-2020
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Cios Alpha Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA20-10308191 and VA30-10308194 The Cios Alpha and Cios Spin are a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
Status
Terminated
Initiated
2020-04-17
Posted
2020-05-04
Terminated
2020-08-25
Reason
The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.
Root cause
Device Design
Action
URGENT: MEDICAL DEVICE CORRECTION notification letters dated 4/17/20 were sent to customers. What is the impact to the operation of the system and what are the possible risks? There is no impact to the operation of the system, however, unplugging the main cable can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector which can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. How was the subject identified and what is the root cause? The issue was identified by the manufacturer during system assembly when the plug of the main cable was unplugged. The root cause is a design flaw in a generator component of the generator Polydoros M25 revision 04. What actions do we recommend urgently to mitigate possible risks? We strongly recommend that you do not unplug the main cable until the generator is repaired by a service engineer. How will the corrective action be implemented and what is the efficiency of the corrective actions? Siemens will correct the hardware error via Update Instruction AX028/20/S in which a Siemens Service Engineer will repair the generator at your site. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the correction, the cause is eliminated, and any recurrence of this potential fault is prevented. Acknowledge Receipt of this Safety Advisory Notice Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein.
Quantity
58
Distribution
US Nationwide Distribution to states of: AR, AZ, CA, FL, IL, IN, MA, MD, MI, MN, NC, NE, NH, NJ, NY, OH, OK, SC, TN, TX, UT, VA, WA, and Puerto Rico.
Lot, serial or model codes
Serial Numbers: 11200 13288 13259 12339 13241 10019 12377 13219 12946 13232 13045 12618 12972 40109 40110 40073 40057 40013 40082 40093 40078 40107 40108 40111 40120 40126 40134 40135 40104 40074 40095 40076 40080 40058 40077 40081 40055 40054 40075 40125 40113 40102 40071 40061 40094 40124 40070 40103 40118 40119 40121 40084 40122 40123 40083 40062 40130 40116
510(k) numbers
["K181560"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register