Baxter Healthcare Corporation: FlowCOUPLER. An implantable device that is used to detect blood flow in vessels. Used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Baxter Healthcare.

Recall number
Z-2126-2014
Recalling firm
Baxter Healthcare Corporation
Product
FlowCOUPLER. An implantable device that is used to detect blood flow in vessels. Used in the end-to-end anastomosis of veins and arteries normally encountered in microsurgical and vascular reconstructive procedures. The FlowCOUPLER System is intended to detect blood flow and confirm vessel patency intra-operatively and post-operatively at the anastomotic site. Baxter Healthcare.
Status
Terminated
Initiated
2014-07-11
Posted
2014-07-29
Terminated
2015-05-27
Reason
Baxter Healthcare is recalling the FlowCOUPLER Device because there is a possibility that some units may not maintain electrical continuity during handling intra-operatively and post-operatively, resulting in a loss of Doppler signal. This may result in the inability to monitor anastomotic patency, with the possibility that blood flow may not be appropriately monitored.
Root cause
Process control
Action
Synovis Micro Companies Alliance, Inc. (a subsidiary of Baxter International Inc.) sent an Urgent Medical Device Recall letter on July 11, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter informs the customers that the recall is being conducted as a precautionary measure due to the possibility that some units may not maintain electrical continuity during handling intra-operatively and post-operatively, resulting in a loss of Doppler signal. This may result in the inability to monitor anastomotic patency, with the possibility that blood flow may not be appropriately monitored. The recall letter informs the customers of the safety risk and and actions to be taken. Customers with questions about the recall are instructed to contact the Synovis MCA Customer Service unit at (800) 510-3318 or (205) 941-0111.
Quantity
1,315 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of United Kingdom, Germany, Sweden, Spain, Israel and Denmark.
Lot, serial or model codes
Lot Number SPSGT14-02B0027a SPSGT14-04D0005 SPSGT14-04D0023  SPSGT13-10J0016 SPSGT13-11K0041 SPSGT13-12L0008 SPSGT14-01A0009  SPSGT14-02B0026 SPSGT14-04D0015 SPSGT14-04D0026  SPSGT13-10J0017 SPSGT13-11K0039 SPSGT13-11K0042 SPSGT13-12L0015 SPSGT14-01A0028  SPSGT14-02B0028 SPSGT14-02B0033 SPSGT14-04D0014 SPSGT14-04D0027 SPSGT14-05E0005  SPSGT13-10J0018 SPSGT13-11K0040 SPSGT13-12L0007 SPSGT13-12L0012 SPSGT14-01A0027   Part No. 5151-01200-010 5151-01200-010 5151-01200-010   5151-01200-011 5151-01200-011 5151-01200-011 5151-01200-011   5151-01250-010 5151-01250-010 5151-01250-010   5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011 5151-01250-011   5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010 5151-01300-010   5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011 5151-01300-011
510(k) numbers
["K132727"]
PMA numbers
not on file
Product code
DPW
Firm FEI
3000121413
Firm city
Westlake Village
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register