Esaote S.p.A.: Magnetic Resonance System G-scan Brio (Part Number /Model No.101002000)

Recall number
Z-2124-2016
Recalling firm
Esaote S.p.A.
Product
Magnetic Resonance System G-scan Brio (Part Number /Model No.101002000)
Status
Terminated
Initiated
2016-06-06
Posted
2016-07-05
Terminated
2017-05-30
Reason
The device is being recalled because dynamic loading forces from larger patients can fracture the extremity portion of the patient table. Additionally, a weight limit is added to the patient seat for weight-bearing examinations.
Root cause
Other
Action
Esaote initiated a voluntary recall of the G-scan Brio Magnetic Resonance System on 06/06/2016 with letters sent via certified, due to a potential weakness on the extremity areas of the patient table that could lead to a stress fracture of the table. The firm has instructed customers to refer to the Medical Device Safety Notification provided with the recall notice, date and sign this letter in the space provided on the last page and send it back via email to quality@esaoteusa.com. If customers have questions regarding the recall, please call 1-317-813-6030. For additional information, please contact us by phone during normal business hours, Monday-Friday from 8:00am 5:00pm Eastern Standard Time, via email at quality@esaoteusa.com, or at any time by visiting our website. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online at www.fda.gov/medwatch/report.htm, by regular mail using the postage-paid, pre-addressed Form FDA3500 available at www.fda.gov/MedWatchgetforms.htm or by fax to 1-800-FDA-0178.
Quantity
11 USA; 29 OUS
Distribution
Domestic: FL, TX, OH, MN, NV; Foreign: No customers in Canada; No VA/DOD.
Lot, serial or model codes
Item No., Unit Description and Serial No.s: 101002000, UNIT, G-SCAN BRIO, 05202; 101002000, UNIT, G-SCAN BRIO, 05203; 101002000, UNIT, G-SCAN BRIO 2.1A; 05205; 101002000, UNIT, G-SCAN BRIO 2.1A, 05207; 101002000 UNIT, G-SCAN BRIO 2.1A, 05208; 101002000 UNIT, G-SCAN BRIO 2.1A, 05209; 101002000 UNIT, G-SCAN BRIO 2.1A, 05210; 101002000 UNIT, G-SCAN BRIO 2.1A, 05211; 101002000 UNIT, G-SCAN BRIO 2.1A, 05305; 101002000 UNIT, G-SCAN BRIO 2.1A, 05308
510(k) numbers
["K142421"]
PMA numbers
not on file
Product code
LNH
Firm FEI
3002806925
Firm city
Firenze
Firm state
not on file
Firm country
Italy

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register